Zimmer Periarticular Locking Plate System Cannulated Drill Standard Quick-Connect 2.7 mm O.D., Item Number/EDI 00236017127, Nonsterile. orthopedic surgical instrument.
Published October 4, 2017 · Recall initiated: May 22, 2017
Summary
FDA records medical device Class II recall Z-3179-2017, published October 4, 2017 and initiated May 22, 2017. Recalling firm: Zimmer Biomet, Inc.; Status: Terminated; Reported quantity: 173 units; Lot: Lot Numbers: 63545430, 63578497, 63586028, 63606286, 63610624, 63635345, 63458048, 63469766, 63475222 & 63530224. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Various trauma and extremity instruments (drill bits and screw taps) were manufactured from raw material with an anomaly. The anomaly may be on or below the surface and can increase the risk of instrument fracture.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Zimmer Biomet, Inc.
- Distributor
- Nationwide in USA; Internationally to: Australia, Brazil, Canada, China, El Salvador, Germany, Guatemala, India, Japan, Panama, Malaysia, Mexico, Netherlands, South Korea, Singapore, Taiwan & Trinidad.
- Lot
- Lot Numbers: 63545430, 63578497, 63586028, 63606286, 63610624, 63635345, 63458048, 63469766, 63475222 & 63530224.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASpinal Rod Cutter, Item number 00392500200. The product is a nonsterile pin cutter - Product Usage: TheDecember 30, 2020
- FDAZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Snap-In O.D. Cup with Spacers, 55November 11, 2020
- FDAZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Neutral, O.D. Cup with Spacers, 55November 11, 2020
- FDAZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, 10 degree inclined face, O.D. CupNovember 11, 2020
- FDAZimmer Periarticular Locking Plate System, Locking Screw Tap Standard Quick-Connect 3.5 mm Diameter, ItemOctober 4, 2017
- FDAZimmer Periarticular Locking Plate System Drill Standard 2.0 mm Diameter, Item Number/EDI 00236017520,October 4, 2017
Questions about this recall
Was Zimmer Periarticular Locking Plate System Cannulated Drill Standard Quick-Connect 2.7 mm O.D., Item Number/EDI 00236017127, Nonsterile. orthopedic surgical instrument. recalled?
Yes. FDA published recall Z-3179-2017 on October 4, 2017. Reported hazard: Various trauma and extremity instruments (drill bits and screw taps) were manufactured from raw material with an anomaly. The anomaly may be on or below the surface and can increase the risk of instrument fracture.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit