STREAMLINE BLOODLINE SET FOR DIALOG DR-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2010M2096A
Published September 25, 2024 · Recall initiated: August 15, 2024
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- B Braun Medical Inc
- Distributor
- US Nationwide distribution.
- Lot
- UDI-DI 04046955674992; All Unexpired Lot Numbers
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States