Captus 4000e Thyroid Uptake System, Models: 1. 5430-30151, 2. 5430-30152, 3. 5430-30154;
Published September 23, 2026 · Recall initiated: July 20, 2026
Summary
FDA records medical device Class II recall Z-3193-2026, published September 23, 2026 and initiated July 20, 2026. Recalling firm: Mirion Technologies (Capintec), Inc.; Status: Ongoing; Reported quantity: 783; Lot: Affected Units: Captus 4000: 1. Model 5430-30151; UDI-DI: 00859942006096; 2. Model 5430-30152; UDI-DI: 00859942006102; 3. Model 5430-30154; UDI-DI: 00859942006423; Devices shipped from April 2019 through February 2024, Serial Numbers 940548 through 941398, that have NOT already replaced the original arm with the redesigned arm. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential for failure of the tension bolt which would allow the spring arm to unexpectedly fall to its lowest position, which is approximately 25 inches (63.5 cm) from the ground.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Mirion Technologies (Capintec), Inc.
- Distributor
- Worldwide distribution - US Nationwide and the countries of ALGERIA, AUSTRIA, BANGLADESH, BRASIL, Brazil, CANADA, CHILE, Costa Rica, Ecuador, France, Hanoi Vietnam, Hong Kong, INDIA, ISRAEL, ITALY, Malaysia, Philippines, POLAND, Saudi Arabia, SOUTH KOREA, Spain, SWEDEN, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIATES, UNITED KINGDOM, VIETNAM.
- Lot
- Affected Units: Captus 4000: 1. Model 5430-30151; UDI-DI: 00859942006096; 2. Model 5430-30152; UDI-DI: 00859942006102; 3. Model 5430-30154; UDI-DI: 00859942006423; Devices shipped from April 2019 through February 2024, Serial Numbers 940548 through 941398, that have NOT already replaced the original arm with the redesigned arm.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAEt Control, Software option for anesthesia gas machine to achieve and maintain targeted end tidal oxygen andSeptember 23, 2026
- FDAAXLE PIN ASSY LARGE ELBOW COBALT CHROME Catalog No. C-M112-2-500 Orthopaedic implant components manufacturedSeptember 23, 2026
- FDACaptus 4000e Thyroid Uptake System. Intended to be used to perform Thyroid Uptakes, Staff Bioassay, and wipeJune 5, 2024
- FDACaptus 4000e Thyroid Uptake System with Well. Intended to be used to perform Thyroid Uptakes, Staff Bioassay,June 5, 2024
- FDACaptus 4000e Thyroid Uptake System with Well and 2 inch shielding. Intended to be used to perform ThyroidJune 5, 2024
- FDACaptus 4000e Thyroid Uptake Systems- Intended to be used by trained Nuclear Medicine Technologists or NuclearFebruary 21, 2024
Questions about this recall
Was Captus 4000e Thyroid Uptake System, Models: 1. 5430-30151, 2. 5430-30152, 3. 5430-30154; recalled?
Yes. FDA published recall Z-3193-2026 on September 23, 2026. Reported hazard: Potential for failure of the tension bolt which would allow the spring arm to unexpectedly fall to its lowest position, which is approximately 25 inches (63.5 cm) from the ground.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail