Ingenia Ambition S - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 1) 782108 2) 781359 3) 782133
Published October 2, 2024 · Recall initiated: September 3, 2024
Summary
FDA records medical device Class II recall Z-3197-2024, published October 2, 2024 and initiated September 3, 2024. Recalling firm: Philips North America Llc; Status: Ongoing; Reported quantity: 64 units; Lot: Model UDI-DI: 782108 (01)00884838098343(21); 781359 (01)00884838090057(21). 782133 N/A Serial Numbers: 27506 27508 27509 27512 27514 27515 27516 27517 27518 27519 27520 27523 27524 27525 27526 27528 27531 48146 48164 48550 48556 48588 48589 48608 48612 48627 48648 48661 48662 48663 48664 48665 48666 48667 48669 48672 48679 48686 48688 48694 48696 48697 48698 48699 48700 48707 48708 48710 48714 48715 48721 48725 48726 48728 48729 48738 48744 48749 48754 48800 48804 48810 48813 48833. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips North America Llc
- Distributor
- Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Iceland, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Maldives, Mauritius, Mexico, Morocco, Netherlands, Pakistan, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Switzerland, Thailand, T®rkiye, United Kingdom, Utd.Arab.Emir., Uzbekistan, Vietnam.
- Lot
- Model UDI-DI: 782108 (01)00884838098343(21); 781359 (01)00884838090057(21). 782133 N/A Serial Numbers: 27506 27508 27509 27512 27514 27515 27516 27517 27518 27519 27520 27523 27524 27525 27526 27528 27531 48146 48164 48550 48556 48588 48589 48608 48612 48627 48648 48661 48662 48663 48664 48665 48666 48667 48669 48672 48679 48686 48688 48694 48696 48697 48698 48699 48700 48707 48708 48710 48714 48715 48721 48725 48726 48728 48729 48738 48744 48749 48754 48800 48804 48810 48813 48833
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPhilips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333October 7, 2026
- FDAAchieva 1.5T, Achieva 1.5T Conversion and Achieva 1.5T Initial system; Model Numbers (REF): 1. 781196 (AchievaSeptember 30, 2026
- FDAIngenia 3.0T and Ingenia 3.0T CX; Model Numbers (REF): 1. 781342(3.0T), 2. 781377(3.0T), 3. 782103(3.0T), 4.September 30, 2026
- FDAIngenia 1.5T, Ingenia 1.5T CX and Ingenia 1.5T S; Model Numbers (REF): 1. 781315(1.5T), 2. 781341(1.5T), 3.September 30, 2026
- FDAIngenia 3.0T CX - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/orOctober 2, 2024
- FDAIngenia Ambition X - For use as a diagnostic device to obtain cross-sectional images, spectroscopic imagesOctober 2, 2024
Questions about this recall
Was Ingenia Ambition S - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 1) 782108 2) 781359 3) 782133 recalled?
Yes. FDA published recall Z-3197-2024 on October 2, 2024. Reported hazard: Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.
What should I do if I have this product?
Voluntary: Firm initiated · Letter