Back to search
FDAFDA · Food and drugs·Recall number Z-3203-2024

SmartPath to dStream for 1.5T - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 1) 782112; 2) 781260

Published October 2, 2024 · Recall initiated: September 3, 2024

Summary

FDA records medical device Class II recall Z-3203-2024, published October 2, 2024 and initiated September 3, 2024. Recalling firm: Philips North America Llc; Status: Ongoing; Reported quantity: 32 units; Lot: Model UDI-DI: 782112 (01)00884838098886(21); 781260 (01)00884838095076(21). Serial Numbers; 8876 10190 10521 13021 13023 18520 18892 18942 20159 21127 21158 21513 21726 21803 22027 22048 22093 22247 30054 30273 32250 32355 32412 32419 32469 32539 32724 32827 33371 33639 39070 75067. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Philips North America Llc
Distributor
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Iceland, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Maldives, Mauritius, Mexico, Morocco, Netherlands, Pakistan, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Switzerland, Thailand, T®rkiye, United Kingdom, Utd.Arab.Emir., Uzbekistan, Vietnam.
Lot
Model UDI-DI: 782112 (01)00884838098886(21); 781260 (01)00884838095076(21). Serial Numbers; 8876 10190 10521 13021 13023 18520 18892 18942 20159 21127 21158 21513 21726 21803 22027 22048 22093 22247 30054 30273 32250 32355 32412 32419 32469 32539 32724 32827 33371 33639 39070 75067
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was SmartPath to dStream for 1.5T - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 1) 782112; 2) 781260 recalled?

Yes. FDA published recall Z-3203-2024 on October 2, 2024. Reported hazard: Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.

What should I do if I have this product?

Voluntary: Firm initiated · Letter