TRIDENT 0¿ POLY INSERT 28mm ID 620-00-28D Orthopaedic implant components manufactured from UHMWPE (polyethylene)
Published September 23, 2026 · Recall initiated: July 30, 2026
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Howmedica Osteonics Corp.
- Distributor
- Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.
- Lot
- 620-00-28D UDI-DI 07613327313581 Lots MHANP2 MHD1N0
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States