Back to search
FDAFDA · Food and drugs·Recall number Z-3208-2026

Convenience kits MINOR PACK DYNJ58421A SPRINGHILL MINOR PACK DYNJ40734F

Published September 23, 2026 · Recall initiated: July 31, 2026

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Kits may exhibit open seals in sterile pouch packaging. Open seal may compromise product sterility. If not detected prior to use, issue may lead to contamination of sterile field or patient infection if product is used.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Medline Industries, LP
Distributor
US Nationwide distribution in states of Arkansas, Florida, North Carolina, Maryland, and Texas.
Lot
DYNJ58421A UDI-DI 10193489970487 Lot 25HBQ049 DYNJ40734F UDI-DI 10198459162084 Lots 25GBO299 25IBS607
Risk classification
Class II
Area (FDA)
device
Country
United States