Percuvance Gripper Grasper Tool Tip The Percutaneous Surgical System with 5mm attachments is indicated for the means to penetrate soft tissue to access certain areas of the abdomen. The system is used to grasp, manipulate, cut, cauterize and deliver Hem-o-lok ligating clips to soft tissue during laparoscopic surgery.
Published October 4, 2017 · Recall initiated: July 19, 2017
Summary
FDA records medical device Class II recall Z-3219-2017, published October 4, 2017 and initiated July 19, 2017. Recalling firm: Teleflex Medical; Status: Terminated; Reported quantity: 20,039 total (US and OUS); Lot: Lot #'s: 73A1600103, 73B1600340, 73C1600290, 73C1600451, 73E1600256, 73G1600559, 73G1600726, 73H1600147, 73H1600340, 73H1600900, 73J1600599, 73K1600501, 73L1600558, and 73M1500034. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
For some of the recalled products, the security of the connection between the shaft and the tool tip may be compromised. For certain other recalled lots of the products, the jaws of the tool tips may close abruptly.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Teleflex Medical
- Distributor
- Worldwide Distribution - US including AL, AZ, CA, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MN, MO, NC, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, WI, and WV Internationally to United Arab Emirates, Vienna, Australia, Belgium, Switzerland, China, Germany, Czech Republic, Spain, Finland, France, United Kingdom, Hong Kong, Ireland, Israel, Iran, Italy, Japan, South Korea, R®publique du Liban, Netherlands, Norway, Qatar, Romania, Saudi Arabia, Republic of Singapore, Turkey, Taiwan, and South Africa
- Lot
- Lot #'s: 73A1600103, 73B1600340, 73C1600290, 73C1600451, 73E1600256, 73G1600559, 73G1600726, 73H1600147, 73H1600340, 73H1600900, 73J1600599, 73K1600501, 73L1600558, and 73M1500034.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAHudson RCI Sheridan EZ-ENDO: a) 5.0 mm, REF 5-22510 b) 5.5 mm, REF 5-22511 c) 7.0 mm, REF 5-22514 d) 7.5 mm,September 18, 2019
- FDAHudson RCI Sheridan/HVT: a) 5.0 mm, REF 5-10310 b) 5.5 mm, REF 5-10311 c) 6.0 mm, REF 5-10312 d) 6.5 mm, REFSeptember 18, 2019
- FDAHudson RCI Sheridan: 1) HVT 6.0 mm, Product Code 5-10312 2) HVT 6.5 mm, Product Code 5-10313 3) HVT 7.0 mm,June 26, 2019
- FDARUSCH LASERTUBE (Rubber), Order No. : 102004-000050, REF 102004, SIZE I.D. mm 5,0June 5, 2019
- FDAUltradent Products, Inc. Paper Points, Size 50 PN #1559, 200 per container, sterile. Product Usage: TheOctober 4, 2017
- FDAPercuvance Maryland Dissector Tool Tip The Percutaneous Surgical System with 5mm attachments is indicated forOctober 4, 2017
Questions about this recall
Was Percuvance Gripper Grasper Tool Tip The Percutaneous Surgical System with 5mm attachments is indicated for the means to penetrate soft tissue to access certain areas of the abdomen. The system is used to grasp, manipulate, cut, cauterize and deliver Hem-o-lok ligating clips to soft tissue during laparoscopic surgery. recalled?
Yes. FDA published recall Z-3219-2017 on October 4, 2017. Reported hazard: For some of the recalled products, the security of the connection between the shaft and the tool tip may be compromised. For certain other recalled lots of the products, the jaws of the tool tips may close abruptly.
What should I do if I have this product?
Voluntary: Firm initiated · Letter