Back to search
FDAFDA · Food and drugs·Recall number Z-3241-2024

ENT, COEN27N; Medical convenience kit

Published October 2, 2024 · Recall initiated: August 7, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
American Contract Systems, Inc.
Distributor
US Nationwide distribution in the states of MO, MN, MA, OH, NE.
Lot
UDI-DI: 00191072206244; Lot: 2405092; Exp: 05/09/2025
Risk classification
Class II
Area (FDA)
device
Country
United States