Back to search
FDAFDA · Food and drugs·Recall number Z-3248-2024

Selenia Dimensions/ 3Dimensions-Digital breast tomosynthesis Software Version Dimensions 1.12.0 and 3Dimensions 2.3.0 system software as follows Model/Catalog Number: 1. ASY-16824 FMI UPGRADE KIT, DIMENSIONS 1.12.0.3 SYSTEM SOFTWARE UPGRADE FOR 1.11.1 SYSTEMS 2. ASY-16825 FMI UPGRADE KIT, 3DIMENSIONS 2.3.0.3 SYSTEM SOFTWARE UPGRADE FOR 2.2.1 SYSTEMS 3. ASY-17219 FMI UPGRADE KIT, DIMENSIONS 1.12.0.3 SYSTEM SOFTWARE UPGRADE FOR 1.10.0 SYSTEMS 4. ASY-17220 FMI UPGRADE KIT, 3DIMENSIONS 2.3.0.3 SYSTEM SOFTWARE UPGRADE FOR 2.1.0 SYSTEMS 5. ASY-17221 FMI UPGRADE KIT, DIMENSIONS 1.12.0.3 SYSTEM SOFTWARE UPGRADE FOR 1.11.0 SYSTEMS 6. ASY-17222 FMI UPGRADE KIT, 3DIMENSIONS 2.3.0.3 SYSTEM SOFTWARE UPGRADE FOR 2.2.0 SYSTEMS 7. 3DM-LIC-SW-UP-V2.3 DIMENSIONS SOFTWARE 2.3 (3DIMENSIONS) 8. SDM-LIC-SW-UP-V1.12 DIMENSION SOFTWARE 1.12 (SELENIA DIMENSIONS)

Published October 2, 2024 · Recall initiated: August 19, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Identified an issue with Selenia Dimensions 1.12.0 and 3Dimensions 2.3.0 system software when used in combination with C-View software in which the quality of the final C-View synthesized 2D image is impacted and may appear blurry and result in misdiagnosis with the possibility being a false negative diagnosis

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Hologic Inc
Distributor
Worldwide distribution - US Nationwide and the countries of Belgium, Canada, Israel, Korea, Republic of, Netherlands.
Lot
UDI-DI: 15420045505636, 15420045511897 Selenia Dimensions 1.12.0 and 3Dimensions 2.3.0 system software when used in combination with C-View¿ software
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card