ADVIA Centaur aTPO Assay, Material Number 10630886 (100 Test) and 10630887 (500 Test)

Published October 9, 2024 · Recall initiated: August 21, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Positive bias in patient results at and below the cut-off of 60 U/mL with the lots listed in the table below when compared to alternate lots on the Atellica IM analyzer.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Siemens Healthcare Diagnostics, Inc.
Distributor
Worldwide distribution - US Nationwide.
Lot
UDI-DI: Material Number 10630886 (100 Test) - 00630414587837; Material Number 10630887 (500 Test) - 00630414587844; Lot Number: 26053337 26054337 26051337 26052337
Risk classification
Class II
Area (FDA)
device
Country
United States