Ingenia 1.5T, 1.5T CX, and 1.5T S; Model number (REF): 1. 781315 (1.5T), 2. 781341 (1.5T), 3. 781396 (1.5T), 4. 782115 (1.5T), 5. 782140 (1.5T), 6. 781262 (1.5T CX), 7. 781347 (1.5T S); Software Versions: R11, R12 and R13;
Published October 7, 2026 · Recall initiated: August 28, 2026
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Philips North America
- Distributor
- Worldwide distribution - Nationwide and the countries of Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Equator. Guinea, Finland, France, French Guiana, Georgia, Germany, Ghana, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kyrgyzstan, Latvia, Lebanon, Luxembourg, Malaysia, Malta, Martinique, Mauritius, Mexico, Moldova, Monaco, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Palestine, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Tunisia, T¿rkiye, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Vietnam.
- Lot
- 1. Model number (REF): 1. 781315 (1.5T); UDI-DI: 00884838099715; Serial Numbers: 61005, 61007, 61014, 61033, 61036, 61047, 61056, 61057, 61076, 61087, 61088, 61105, 61116; 2. Model Number (REF): 781341 (1.5T); UDI-DI: 00884838055322; Serial Numbers: 70042, 70051, 70056, 70062, 70203, 70227, 70229, 70249, 70257, 70262, 70285, 70300, 70311, 70383, 70424, 70432, 70438, 70445, 70453, 70460, 70472, 70588, 70598, 70619, 70638, 70642, 70645, 70647, 70708, 70717, 70724, 70728, 70743, 70750, 70751, 70787, 70801, 70821, 70825, 70836, 70846, 70875, 70876, 70886, 70889, 70896, 70904, 70920, 70957, 70961, 70967, 70970, 70979, 70988, 84049, 84076, 84079, 84080, 84141, 84162, 84170, 84173, 84186, 84205, 84212, 84219, 84233, 84250, 84254, 84266, 84267, 84281, 84289, 84329, 84332, 84335, 84343, 84359, 84360, 84368, 84379, 84388, 84392, 84403, 84415, 84431, 84433, 84436, 84450, 84451, 84467, 84476, 84491, 84500, 84510, 84539, 84549, 84555, 84556, 84568, 84596, 84634, 84635, 84641, 84642, 84652, 84660, 84708, 84713, 84714, 84721, 84741, 84745, 84768, 84804, 84839, 84840, 84853, 84858, 84867, 84869, 84893, 84904, 84908, 84978, 84987, 87068, 87076, 87092, 87196, 87197, 87259; 3. Model Number (REF): 781396 (1.5T); UDI-DI: 00884838009820; Serial Numbers: 41000, 41001, 41021, 41034, 41046, 41056, 41061, 41091, 41116, 41145, 41194, 41231, 41249, 41253, 41254, 41255, 41352, 41354, 41356, 41370, 41400, 41408, 41422, 41428, 41456, 41475, 41492, 41513, 41515, 41548; 4. Model Number (REF): 782115 (1.5T); UDI-DI: 00884838099043; Serial Numbers: 61186, 61191, 61196, 61219, 61228, 61236, 61238, 61272, 61276, 61317, 61382, 61383, 61384, 61386, 61387, 61388, 61390, 61396, 61399, 61401, 61402, 61408, 61411, 61418, 61430; 5. Model Number (REF): 782140 (1.5T); UDI-DI: 00884838108646; Serial Numbers: 45608, 45609; 6. Model Number (REF): 781262 (1.5T CX); UDI-DI: 00884838068445; Serial Numbers: 77047, 77058, 77059, 77120, 77143, 77151, 77158, 77161, 77179, 77180, 77182; 7. Model Number (REF): 781347 (1.5T S); UDI-DI: 00884838068421; Serial Numbers: 81027, 81033, 81145;
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASmartPath to dStream for 1.5T, SmartPath to dStream for 3.0T, SmartPath to dStream for XR and 3.0T, and Smar…October 7, 2026
- FDAUpgrade dStream; Model number (REF): 782127; Software Versions: R11, R12 and R13;October 7, 2026
- FDAMR 7700; Model number (REF): 1. 782120, 2. 782153, 3. 782155, 4. 782199; Software Versions: R11, R12 and R1…October 7, 2026
- FDAUpgrade to MR 7700; Model number (REF): 782130; Software Versions: R11, R12 and R13;October 7, 2026
- FDAIngenia Blue Seal SE and Blue Seal XE; Model number (REF): 1. 782185 (SE), 2. 782192 (XE); Software Version…October 7, 2026
- FDAIngenia 3.0T and 3.0T CX; Model number (REF): 1. 781342 (3.0T), 2. 781377 (3.0T), 3. 782103 (3.0T), 4. 78127…October 7, 2026
- FDANOxBOXi Nitric Oxide Delivery System. Model Number: NOXBOX-I.October 7, 2026
- FDAIngenia Ambition S and Ingenia Ambition X.; Model number (REF): 1. 781359 (S), 2. 782108 (S), 3. 782139 (S),…October 7, 2026
Questions about this recall
Was Ingenia 1.5T, 1.5T CX, and 1.5T S; Model number (REF): 1. 781315 (1.5T), 2. 781341 (1.5T), 3. 781396 (1.5T), 4. 782115 (1.5T), 5. 782140 (1.5T), 6. 781262 (1.5T CX), 7. 781347 (1.5T S); Software Versions: R11, R12 and R13; recalled?
Yes. FDA published recall Z-3308-2026 on October 7, 2026. Reported hazard: When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.
What should I do if I have this product?
Voluntary: Firm initiated · Letter
Sell products like this?
Recalert watches your whole catalog for recalls like this one, automatically.
Recalert also tracks expiration dates so you can sell stock before it becomes a loss.