FDAFDA · Food and drugs·Recall number D-1383-2012
LifeGas OXYGEN COMPRESSED UN1072 USP, Distributed by: Linde
Published June 27, 2012 · Recall initiated: May 4, 2012
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
CGMP Deviations: The pressure gages, vacuum gages, and thermometer had surpassed the calibration expiry period, which may have resulted in overfill/underfill of oxygen cylinders.
What to do
Voluntary: Firm initiated · Other
Recall details
- Manufacturer
- Linde Gas LLC
- Distributor
- Distributed only within Ohio. No Federal Government, foreign or Canadian consignees received product.
- Lot
- LOT#: 117H201 LOT#: 117H202
- Risk classification
- Class II
- Area (FDA)
- drug
- Country
- United States
Sell products like this?
Recalert watches your whole catalog for recalls like this one, automatically.