MiniMed Insulin Pump Device. MiniMed 670G, REF: MMT-1580, MMT-1581, MMT-1582, MMT-1780, MMT-1781, MMT-1782, MMT-1740, MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762; MiniMed 770G, REF: MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892; MiniMed 780G, REF: MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896
Published October 9, 2024 · Recall initiated: July 31, 2024
Summary
FDA records medical device Class I recall Z-0003-2025, published October 9, 2024 and initiated July 31, 2024. Recalling firm: Medtronic MiniMed, Inc.; Status: Ongoing; Reported quantity: 526,558; Lot: REF/UDI-DI(GTIN): MMT-1780/00763000190453, 00643169946057, 00763000365882, 00763000096984, 00763000283445, 763000283513, 763000283506, 763000365882, 763000283520, 76300028351303, 763000393786, 763000166526, 76300028351301, 643169939202, 763000190460, 763000090203, 763000256821, 643169939219, 763000072537, 763000365912, 00763000090203, 00763000190460, 00763000283520, 00763000393786; MMT-1781/00763000187415, 00763000283568, 00763000283544, 00763000404536, 00763000283551, 00763000283537, 00763000365899, 00763000371937, 00763000225575, 00763000225582, 00763000235765, 00763000187378, 00763000140601, 763000240707, 763000187378, 763000140601, 763000283544; MMT-1782/00763000187422, 00763000235772, 00763000283599, 00763000264727, 00763000404543, 00763000283582, 00763000371944, 00763000365905, 00763000283575, 00643169957640, 00763000157463, 00763000225612, 763000225629, 763000225612, 763000264727; MMT-1741/00763000408916, 00763000408930; MMT-1742/00763000408947, 00763000250232, 00763000408954; MMT-1760/763000278427, 763000283445, 643169946057, 763000190453, 763000096984, 643169946040, 00643169946057, 00763000190453, 00763000283445, 763000367084, 763000179649, 763000395858, 00763000096984; MMT-1761/763000187392, 00763000165451, 00763000165420, 00763000165468, 00763000165444, 00763000165437, 00763000257309, 00763000165413, 00763000289065, 00763000289089, 00763000289041, 00763000367657, 00763000289195, 00763000289096, 00763000289072, 00763000289058, 00763000324797, 00763000374518, 00763000374532, 00763000374525, 00763000374556, 00763000399849, 00763000399856, 00763000374587, 763000367657; MMT-1762/763000264710, 00763000264710, 00763000165475, 00763000165482, 00763000165499, 00763000165505, 00763000165512, 00763000165529, 00763000165536, 00763000165543, 00763000165550, 00763000203313, 00763000216481, 00763000252076, 00763000283476, 00763000289102, 00763000289119, 00763000289126, 00763000289133, 00763000289140, 00763000289164, 00763000289171, 00763000289188, 00763000324780, 00763000374594, 00763000374600, 00763000374624, 00763000399986; MMT-1880/00763000414344, 00763000545604, 00763000439859, 763000578640, 763000586195, 76300058619502, 763000439866, 763000413750, 76300057864002, 763000414344, 763000545604, 763000640095, 763000439859, 76300058619501, 00763000413750, 00763000586188; MMT-1881/00763000416546, 00763000439903, 00763000629892, 763000416546; MMT-1882/00763000416553, 00763000439910, 00763000545505, 00763000629908; MMT-1890/763000545604, 763000414344, 763000439859, 763000640095, 00763000414344, 00763000439859, 00763000545604, 76300054560402, 00763000640095, 76300064009502; MMT-1891/00763000642242, 00763000456634, 00763000642235, 00763000418151, 00763000482596; MMT-1892/00763000416553, 00763000439910, 00763000545505, 00763000629908; MMT-1884/00763000414344, 76300081937801, 76300081938501, 763000819378, 763000819385, 00763000586195, A7630007585401, 00763000758547, 00763000439866, 00763000578640, 00763000439859, 00763000545604, 00763000640095, 00763000819392, 763000545604, 00763000758530, 00763000504762; MMT-1885/00763000384258, 00763000521523, 00763000442187, 00763000521509, 00763000578671, 00763000639228, 00763000421939, 00763000597900, 00763000854942, 00763000854829, A7630006392201, 763000384272, 763000521523, 763000854829; MMT-1886/00763000384265, 00763000441449, 00763000521530, 00763000521516, 00763000578688, 00763000442194, 00763000639235, 763000384289, 763000521530, 76300044240801, 763000688165, 76300063923501; MMT-1894/76300086566501, 76300075854701, 763000758547, 763000819392, 763000865665, 00763000758547, 00763000819392; MMT-1895/00763000411411, 00763000734435, 00763000411473, 00763000411466, 00763000541569, 00763000411480, 00763000420703, 00763000411428, 00763000596217, 00763000411404, 00763000533861, 00763000411381, 00763000411503, 00763000596187, 00763000411398, 00763000596194, 00763000411374, 00763000411442, 00763000411435, 00763000411497, 00763000411459, 00763000596224, 00763000411633, 00763000596231, 00. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overall, it shortens the time between low battery alarms and increases the frequency with which the user must insert a new battery. Further, the defect may also deliver the low battery alarm when there is significantly less time before the battery completely runs out, from the expected up to 10-hour buffer the low alert is supposed to signal, leading to device powering down and stopping insulin delivery unexpectedly. Per the firm, a subset of pumps with the defect showed the low alert alarm was delivered ~2.5 hours before the device shut off, however, patient reports indicate the time could be even shorter. Replacing the battery will not resolve this issue and the short battery life and the truncated low battery alert lead time will continue to occur. Early battery depletion could result in power loss and insulin delivery could be interrupted, resulting in under delivery of insulin potentially leading to hyperglycemia and/or DKA.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Medtronic MiniMed, Inc.
- Distributor
- US: CT, MI, PA, WA, IA, NY, ND, AZ, TX, OH, NC, AL, MN, IN, NJ, KY, UT, CA, FL, VA, MS, NM, NV, TN, GA, MA, NH, OK, VT, IL, ME, SC, LA, WY, RI, SD, KS, WI, MD, CO, DE, AR, AK, ID, MO, NE, WV, MT, OR, DC, HI, VI, PR. OUS: Worldwide
- Lot
- REF/UDI-DI(GTIN): MMT-1780/00763000190453, 00643169946057, 00763000365882, 00763000096984, 00763000283445, 763000283513, 763000283506, 763000365882, 763000283520, 76300028351303, 763000393786, 763000166526, 76300028351301, 643169939202, 763000190460, 763000090203, 763000256821, 643169939219, 763000072537, 763000365912, 00763000090203, 00763000190460, 00763000283520, 00763000393786; MMT-1781/00763000187415, 00763000283568, 00763000283544, 00763000404536, 00763000283551, 00763000283537, 00763000365899, 00763000371937, 00763000225575, 00763000225582, 00763000235765, 00763000187378, 00763000140601, 763000240707, 763000187378, 763000140601, 763000283544; MMT-1782/00763000187422, 00763000235772, 00763000283599, 00763000264727, 00763000404543, 00763000283582, 00763000371944, 00763000365905, 00763000283575, 00643169957640, 00763000157463, 00763000225612, 763000225629, 763000225612, 763000264727; MMT-1741/00763000408916, 00763000408930; MMT-1742/00763000408947, 00763000250232, 00763000408954; MMT-1760/763000278427, 763000283445, 643169946057, 763000190453, 763000096984, 643169946040, 00643169946057, 00763000190453, 00763000283445, 763000367084, 763000179649, 763000395858, 00763000096984; MMT-1761/763000187392, 00763000165451, 00763000165420, 00763000165468, 00763000165444, 00763000165437, 00763000257309, 00763000165413, 00763000289065, 00763000289089, 00763000289041, 00763000367657, 00763000289195, 00763000289096, 00763000289072, 00763000289058, 00763000324797, 00763000374518, 00763000374532, 00763000374525, 00763000374556, 00763000399849, 00763000399856, 00763000374587, 763000367657; MMT-1762/763000264710, 00763000264710, 00763000165475, 00763000165482, 00763000165499, 00763000165505, 00763000165512, 00763000165529, 00763000165536, 00763000165543, 00763000165550, 00763000203313, 00763000216481, 00763000252076, 00763000283476, 00763000289102, 00763000289119, 00763000289126, 00763000289133, 00763000289140, 00763000289164, 00763000289171, 00763000289188, 00763000324780, 00763000374594, 00763000374600, 00763000374624, 00763000399986; MMT-1880/00763000414344, 00763000545604, 00763000439859, 763000578640, 763000586195, 76300058619502, 763000439866, 763000413750, 76300057864002, 763000414344, 763000545604, 763000640095, 763000439859, 76300058619501, 00763000413750, 00763000586188; MMT-1881/00763000416546, 00763000439903, 00763000629892, 763000416546; MMT-1882/00763000416553, 00763000439910, 00763000545505, 00763000629908; MMT-1890/763000545604, 763000414344, 763000439859, 763000640095, 00763000414344, 00763000439859, 00763000545604, 76300054560402, 00763000640095, 76300064009502; MMT-1891/00763000642242, 00763000456634, 00763000642235, 00763000418151, 00763000482596; MMT-1892/00763000416553, 00763000439910, 00763000545505, 00763000629908; MMT-1884/00763000414344, 76300081937801, 76300081938501, 763000819378, 763000819385, 00763000586195, A7630007585401, 00763000758547, 00763000439866, 00763000578640, 00763000439859, 00763000545604, 00763000640095, 00763000819392, 763000545604, 00763000758530, 00763000504762; MMT-1885/00763000384258, 00763000521523, 00763000442187, 00763000521509, 00763000578671, 00763000639228, 00763000421939, 00763000597900, 00763000854942, 00763000854829, A7630006392201, 763000384272, 763000521523, 763000854829; MMT-1886/00763000384265, 00763000441449, 00763000521530, 00763000521516, 00763000578688, 00763000442194, 00763000639235, 763000384289, 763000521530, 76300044240801, 763000688165, 76300063923501; MMT-1894/76300086566501, 76300075854701, 763000758547, 763000819392, 763000865665, 00763000758547, 00763000819392; MMT-1895/00763000411411, 00763000734435, 00763000411473, 00763000411466, 00763000541569, 00763000411480, 00763000420703, 00763000411428, 00763000596217, 00763000411404, 00763000533861, 00763000411381, 00763000411503, 00763000596187, 00763000411398, 00763000596194, 00763000411374, 00763000411442, 00763000411435, 00763000411497, 00763000411459, 00763000596224, 00763000411633, 00763000596231, 00
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAGeneric Name: MiniMed Go App Generic Name: MiniMed Go App Model/Catalog Number: MMT-8600 (iOS), MMT-8601October 28, 2026
- FDACareLink Clinic, REF: MMT-7350October 28, 2026
- FDABrand Name: MiniMed Flex Product Name: MiniMed Flex Insulin Pump Model/Catalog Numbers: MMT-8062; MMT-8072October 14, 2026
- FDAMiniMed 780G Insulin Pump, Model: MMT-1885CLW, MMT-1886CLWAugust 26, 2026
- FDAMiniMed 620G Insulin Pump, REF: MMT-1510, MMT-1710, MMT-1750; MiniMed 630G Insulin Pump, REF: MMT-1515,October 9, 2024
- FDANeo-Tee T-Piece Resuscitator. Part Numbers 1050805, 1050832. For pediatric emergency respiratory support.October 9, 2024
Questions about this recall
Was MiniMed Insulin Pump Device. MiniMed 670G, REF: MMT-1580, MMT-1581, MMT-1582, MMT-1780, MMT-1781, MMT-1782, MMT-1740, MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762; MiniMed 770G, REF: MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892; MiniMed 780G, REF: MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896 recalled?
Yes. FDA published recall Z-0003-2025 on October 9, 2024. Reported hazard: Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overall, it shortens the time between low battery alarms and increases the frequency with which the user must insert a new battery. Further, the defect may also deliver the low battery alarm when there is significantly less time before the battery completely runs out, from the expected up to 10-hour buffer the low alert is supposed to signal, leading to device powering down and stopping insulin delivery unexpectedly. Per the firm, a subset of pumps with the defect showed the low alert alarm was delivered ~2.5 hours before the device shut off, however, patient reports indicate the time could be even shorter. Replacing the battery will not resolve this issue and the short battery life and the truncated low battery alert lead time will continue to occur. Early battery depletion could result in power loss and insulin delivery could be interrupted, resulting in under delivery of insulin potentially leading to hyperglycemia and/or DKA.
What should I do if I have this product?
Voluntary: Firm initiated · Letter