Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin Pump Model/Catalog Numbers: MMT-8062; MMT-8072 Product Description: The MiniMed Flex pump is an alternate controller enabled (ACE) pump intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. It can reliably and securely communicate with compatible digitally connected devices, including an integrated continuous glucose monitor (iCGM), interoperable Medtronic continuous glucose monitor (CGM), and interoperable automated glycemic controller (iAGC). The pump is intended to be used in conjunction with compatible, digitally connected medical devices for the purpose of drug delivery. The primary user interface and primary display for the MiniMed Flex pump is the MiniMed app, also the user interface for the compatible MiniMed iAGCs, which runs on a compatible iOS or Android device.
Published October 14, 2026 · Recall initiated: August 20, 2026
Summary
FDA records medical device Class II recall Z-0088-2027, published October 14, 2026 and initiated August 20, 2026. Recalling firm: Medtronic MiniMed, Inc.; Status: Ongoing; Reported quantity: 25; Lot: MMT-8072 Lot Code: Serial numbers: FB00015839, FB00016511, FB00014455, FB00011553, FB00015700, FB00011357, FB00019307, FB00014766, FB00014822, FB00015828, FB00018580, FB00010465, FB00014472, FB00015793, FB00015794, FB00015796, FB00015799, FB00015801, FB00015977, FB00017271. UDI-DI/GTIN: 00763000848156 Note: Initially 25 pumps were identified as put into service, however 2 were shipped to Marketing Team Members and recovered. 3 Pumps were returned through complaint process. Final total of pumps available in the field is 20. MMT-8062 Lot Code: Serial numbers: FB00011770, FB00015162, FB00016429, FB00014760, FB00015797 UDI-DI/GTIN: 00763000848149 (Serial Numbers: FB00017271, FB00011770, FB00015162, FB00016429, FB00014760 were returned via complaint process or shipped to Marketing Team Members). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed verification test results in one or more of 9 functional tests. The unverified/failed pumps released to users potentially have one or more of the following issues: inoperable pump at setup, lack of protection for moisture exposure, slow battery charging, disconnection issues with other system components (i.e., continuous glucose monitoring sensor or MiniMed mobile app), reduced ability to detect occlusion events, compromised audio for pump alarms/alerts, and/or inaccurate insulin delivery.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Medtronic MiniMed, Inc.
- Distributor
- U.S. Nationwide distribution in the states of AR, AZ, CO, FL, KS, KY, LA, MI, MN, NJ, NY, OH, PA, SD, TN, TX, and WV.
- Lot
- MMT-8072 Lot Code: Serial numbers: FB00015839, FB00016511, FB00014455, FB00011553, FB00015700, FB00011357, FB00019307, FB00014766, FB00014822, FB00015828, FB00018580, FB00010465, FB00014472, FB00015793, FB00015794, FB00015796, FB00015799, FB00015801, FB00015977, FB00017271. UDI-DI/GTIN: 00763000848156 Note: Initially 25 pumps were identified as put into service, however 2 were shipped to Marketing Team Members and recovered. 3 Pumps were returned through complaint process. Final total of pumps available in the field is 20. MMT-8062 Lot Code: Serial numbers: FB00011770, FB00015162, FB00016429, FB00014760, FB00015797 UDI-DI/GTIN: 00763000848149 (Serial Numbers: FB00017271, FB00011770, FB00015162, FB00016429, FB00014760 were returned via complaint process or shipped to Marketing Team Members).
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAGeneric Name: MiniMed Go App Generic Name: MiniMed Go App Model/Catalog Number: MMT-8600 (iOS), MMT-8601October 28, 2026
- FDACareLink Clinic, REF: MMT-7350October 28, 2026
- FDARollator, Bariatric, Basic Steel, Blue, mechanical walker with four legs and a metal frame intended forOctober 14, 2026
- FDAProduct Name: Nihon Kohden CNS-6200 Series Central Nurse Station, Model/Catalog Number: CN-6201 SoftwareOctober 14, 2026
- FDAMiniMed 780G Insulin Pump, Model: MMT-1885CLW, MMT-1886CLWAugust 26, 2026
- FDAMiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782;August 26, 2026
Questions about this recall
Was Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin Pump Model/Catalog Numbers: MMT-8062; MMT-8072 Product Description: The MiniMed Flex pump is an alternate controller enabled (ACE) pump intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. It can reliably and securely communicate with compatible digitally connected devices, including an integrated continuous glucose monitor (iCGM), interoperable Medtronic continuous glucose monitor (CGM), and interoperable automated glycemic controller (iAGC). The pump is intended to be used in conjunction with compatible, digitally connected medical devices for the purpose of drug delivery. The primary user interface and primary display for the MiniMed Flex pump is the MiniMed app, also the user interface for the compatible MiniMed iAGCs, which runs on a compatible iOS or Android device. recalled?
Yes. FDA published recall Z-0088-2027 on October 14, 2026. Reported hazard: Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed verification test results in one or more of 9 functional tests. The unverified/failed pumps released to users potentially have one or more of the following issues: inoperable pump at setup, lack of protection for moisture exposure, slow battery charging, disconnection issues with other system components (i.e., continuous glucose monitoring sensor or MiniMed mobile app), reduced ability to detect occlusion events, compromised audio for pump alarms/alerts, and/or inaccurate insulin delivery.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit