Ingenuity Core Model Number 728321 Product Usage: Computed Tomography X-ray systems intended to product cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
Published October 12, 2016 · Recall initiated: August 24, 2016
Summary
FDA records medical device Class II recall Z-0010-2017, published October 12, 2016 and initiated August 24, 2016. Recalling firm: Philips Medical Systems; Status: Terminated; Reported quantity: 358; Lot: 30041, 31001, 33084, 52000, 52004, 52007, 52008, 52010, 52012, 52013, 52014, 52015, 52018, 52019, 52020, 52021, 52022, 52024, 52025, 52026, 52027, 52029, 52030, 52031, 52032, 52033, 52034, 52035, 52036, 52037, 52038, 52039, 52040, 52041, 52042, 52043, 52044, 52045, 52046, 52047, 52048, 52050, 25052, 52053, 52054, 52055, 52056, 52057, 52058, 52059, 52061, 52063, 52064, 52065, 52067, 52068, 52069, 52070, 52071, 52072, 52073, 52074, 52075, 52076, 52077, 52078, 52079, 52080, 52081, 52082, 52083, 52084, 132164, 132465, 310001, 310002, 310003, 310009, 310017, 310019, 310022, 310023, 310026, 310032, 310039, 310042, 310052, 310054, 310055, 310059, 310064, 310069, 310070, 310071, 310081, 310087, 310088, 310090, 310096, 310098, 310105, 310107, 310111, 310113, 310116, 310119, 310121, 310125, 310129, 310133, 310134, 310135, 310141, 310143, 310149, 310150, 310151, 310157, 310159, 310160, 310170, 310203, 310204, 310205, 310206, 310207, 310208, 310209, 310210, 310211, 310212, 310213, 310214, 310215, 310216, 310217, 310218, 310219, 310220, 310221, 310222, 310223, 310224, 310225, 310226, 310227, 310228, 310229, 310230, 310231, 310232, 310233, 310234, 310235, 310236, 310237, 310238, 310239, 310240, 310241, 310242, 310243, 310244, 310245, 310246, 310247, 310248, 310249, 310250, 310251, 310252, 310253, 310254, 310255, 310256, 310257, 310258, 310259, 310260, 310261, 310262, 310263, 310264, 310265, 310266, 310267, 310268, 310269, 310270, 310271, 310272, 310273, 310274, 310275, 310276, 310277, 310278, 310279, 310280, 310281, 310282, 310283, 310284, 310285, 310286, 310287, 310288, 310289, 310290, 310291, 310292, 310293, 310294, 310295, 310296, 310297, 310298, 310299, 310300, 310301, 310302, 310303, 310304, 310305, 310306, 310307, 310308, 310309, 310310, 310311, 310313, 310315, 310316, 310317, 310318, 310319, 310320, 310321, 310322, 310323, 310324, 310325, 310326, 310327, 310328, 310329, 310330, 310331, 333002, 333005, 333009, 333010, 333014, 333015, 333018, 333019, 333021, 333030, 333031, 333037, 333041, 333042, 333043, 333044, 333045, 333046, 333048, 333049, 333050, 333051, 333052, 333053, 333054, 333055, 333056, 333059, 333060, 333061, 333062, 333063, 333064, 333065, 333066, 333067, 333068, 333069, 333070, 333071, 333072, 333073, 333074, 333075, 333076, 333077, 333078, 333079, 333080, 333081, 333082, 333083, 333085, 333086, 333087, 333088, 333089, 333090, 333091, 333092, 333093, 333094, 333095, 333096, 333097, 333098, 333099, 333100, 333101, 333102, 333103, 333104, 333105, 333106, 333107, 333108, 333109, 333110, 333111, 333112, 333113, 333114, 333115, 333116, 333117, 333118, 333119, 333120, 333121, 333122, 333123, 333124, 333125, 333126, 333127, 333129, 333130, 333131, 333132, 333133, 333134, 333135, 333136, 333137, 333140, 333144, 333145, 333146, 333150, 333151. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The infant performance test did not meet the test specification for the Infant Head Uniformity test with High Resolution. As a result, when the user performs High Resolution head scans, the reconstructed images may exhibit degraded image quality manifested as non-uniformity of the image, CT number shift greater than 5 Hounsfield units, and reduction in gray/white matter differentiation. The issue is most sever in the High Resolution mode, and the factory reference pediatric scan protocols are set to High Resolution default. Adults head scans have a factory reference protocol that defaults to standard resolution which doesn't experience the issue.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Medical Systems
- Distributor
- Worldwide Distribution - US Nationwide in the states of:: AK, AL, AR, AZ, CA, CO, CT, DE, GA, IA, IL, IN, FL, KS, KY, LA, MA, MD, ME, MI, MO, MS, MT, NC, ND, NE, NJ, NH, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, & WY. and the countries of Algeria, Argentina, Australia, Bangladesh, Belgium, Bolivia, Brazil, Cambodia, China, COlumbia, Cuba, Denmark, Dijbouti, Dominican REpublic, Egypt, Estonia, Finland, France, Germany, Gilbraltar, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Libya, Lithuania, Malaysia, Martinique, Monaco, Mongolia, Mozambique, Myanmar, Netherlands, New Zealand, Norway, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Russian Federation, Senegal, Singapore, Slovakia, South Africa, Spain, Sweden, Seitzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Uniter Arab Emirates, United Kingdom, Uzbekistan, and Vietnam.
- Lot
- 30041, 31001, 33084, 52000, 52004, 52007, 52008, 52010, 52012, 52013, 52014, 52015, 52018, 52019, 52020, 52021, 52022, 52024, 52025, 52026, 52027, 52029, 52030, 52031, 52032, 52033, 52034, 52035, 52036, 52037, 52038, 52039, 52040, 52041, 52042, 52043, 52044, 52045, 52046, 52047, 52048, 52050, 25052, 52053, 52054, 52055, 52056, 52057, 52058, 52059, 52061, 52063, 52064, 52065, 52067, 52068, 52069, 52070, 52071, 52072, 52073, 52074, 52075, 52076, 52077, 52078, 52079, 52080, 52081, 52082, 52083, 52084, 132164, 132465, 310001, 310002, 310003, 310009, 310017, 310019, 310022, 310023, 310026, 310032, 310039, 310042, 310052, 310054, 310055, 310059, 310064, 310069, 310070, 310071, 310081, 310087, 310088, 310090, 310096, 310098, 310105, 310107, 310111, 310113, 310116, 310119, 310121, 310125, 310129, 310133, 310134, 310135, 310141, 310143, 310149, 310150, 310151, 310157, 310159, 310160, 310170, 310203, 310204, 310205, 310206, 310207, 310208, 310209, 310210, 310211, 310212, 310213, 310214, 310215, 310216, 310217, 310218, 310219, 310220, 310221, 310222, 310223, 310224, 310225, 310226, 310227, 310228, 310229, 310230, 310231, 310232, 310233, 310234, 310235, 310236, 310237, 310238, 310239, 310240, 310241, 310242, 310243, 310244, 310245, 310246, 310247, 310248, 310249, 310250, 310251, 310252, 310253, 310254, 310255, 310256, 310257, 310258, 310259, 310260, 310261, 310262, 310263, 310264, 310265, 310266, 310267, 310268, 310269, 310270, 310271, 310272, 310273, 310274, 310275, 310276, 310277, 310278, 310279, 310280, 310281, 310282, 310283, 310284, 310285, 310286, 310287, 310288, 310289, 310290, 310291, 310292, 310293, 310294, 310295, 310296, 310297, 310298, 310299, 310300, 310301, 310302, 310303, 310304, 310305, 310306, 310307, 310308, 310309, 310310, 310311, 310313, 310315, 310316, 310317, 310318, 310319, 310320, 310321, 310322, 310323, 310324, 310325, 310326, 310327, 310328, 310329, 310330, 310331, 333002, 333005, 333009, 333010, 333014, 333015, 333018, 333019, 333021, 333030, 333031, 333037, 333041, 333042, 333043, 333044, 333045, 333046, 333048, 333049, 333050, 333051, 333052, 333053, 333054, 333055, 333056, 333059, 333060, 333061, 333062, 333063, 333064, 333065, 333066, 333067, 333068, 333069, 333070, 333071, 333072, 333073, 333074, 333075, 333076, 333077, 333078, 333079, 333080, 333081, 333082, 333083, 333085, 333086, 333087, 333088, 333089, 333090, 333091, 333092, 333093, 333094, 333095, 333096, 333097, 333098, 333099, 333100, 333101, 333102, 333103, 333104, 333105, 333106, 333107, 333108, 333109, 333110, 333111, 333112, 333113, 333114, 333115, 333116, 333117, 333118, 333119, 333120, 333121, 333122, 333123, 333124, 333125, 333126, 333127, 333129, 333130, 333131, 333132, 333133, 333134, 333135, 333136, 333137, 333140, 333144, 333145, 333146, 333150, 333151
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAIngenuity CT Model Number 728326 Product Usage: Computed Tomography X-ray systems intended to productOctober 12, 2016
- FDA0.2 Micron Filter, 50 mm Product Usage: The 0.2 Micron Filter is a bacteria and particulate filter for aqueousOctober 12, 2016
- FDADiamond of California® Macadamia Halves and Pieces 2.25oz packagesOctober 12, 2016
- FDAMultix Fusion Stationary X-Ray System Product Usage: The Multix Fusion system is a radiographic system used inOctober 12, 2016
- FDADiet Dr Pepper, 12 oz. can/24 pack, UPC 078000083132; Diet Dr Pepper, 12 oz. can/6 pack, UPC N/A; Diet DrOctober 12, 2016
- FDAPanoramic X-ray Model PC-1000, Laser-1000 packed in a crate Product Usage: The PC-1000 will enable the user toOctober 12, 2016
Questions about this recall
Was Ingenuity Core Model Number 728321 Product Usage: Computed Tomography X-ray systems intended to product cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. recalled?
Yes. FDA published recall Z-0010-2017 on October 12, 2016. Reported hazard: The infant performance test did not meet the test specification for the Infant Head Uniformity test with High Resolution. As a result, when the user performs High Resolution head scans, the reconstructed images may exhibit degraded image quality manifested as non-uniformity of the image, CT number shift greater than 5 Hounsfield units, and reduction in gray/white matter differentiation. The issue is most sever in the High Resolution mode, and the factory reference pediatric scan protocols are set to High Resolution default. Adults head scans have a factory reference protocol that defaults to standard resolution which doesn't experience the issue.
What should I do if I have this product?
Voluntary: Firm initiated · Letter