Allura Xper FD20/15; Model Numbers: 722058; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;

Published October 8, 2025 · Recall initiated: September 3, 2025

Summary

FDA records medical device Class II recall Z-0010-2026, published October 8, 2025 and initiated September 3, 2025. Recalling firm: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.; Status: Ongoing; Reported quantity: 300 units (69 US, 231 OUS); Lot: Model Numbers: 722058; UDI-DI: 00884838059115; Serial Numbers: 337, 178, 249, 22, 202, 279, 254, 285, 71, 186, 82, 85, 162, 138, 24, 365, 203, 167, 146, 168, 304, 172, 200, 123, 129, 149, 259, 4, 197, 134, 38, 290, 171, 56, 264, 238, 199, 61, 156, 99, 65, 59, 109, 100, 305, 319, 215, 278, 124, 196, 405, 325, 265, 378, 277, 274, 94, 182, 366, 340, 190, 219, 373, 239, 23, 205, 257, 212, 111, 110, 399, 72, 332, 92, 142, 67, 106, 335, 360, 415, 11, 114, 151, 93, 232, 395, 292, 380, 183, 349, 261, 15, 375, 161, 78, 368, 9, 372, 288, 252, 320, 213, 293, 389, 169, 97, 336, 25, 143, 133, 192, 49, 412, 201, 89, 66, 163, 170, 77, 273, 35, 222, 229, 248, 334, 384, 40, 268, 240, 258, 289, 269, 226, 244, 48, 141, 391, 241, 164, 381, 281, 154, 246, 127, 315, 256, 233, 348, 204, 396, 7, 185, 95, 131, 377, 60, 68, 206, 101, 150, 333, 84, 398, 313, 3, 39, 344, 376, 382, 370, 387, 419, 394, 407, 413, 392, 411, 358, 409, 385, 416, 367, 379, 331, 418, 356, 421, 369, 420, 397, 357, 408, 393, 388, 108, 211, 126, 125, 301, 12, 30, 355, 354, 159, 139, 128, 81, 58, 96, 343, 177, 115, 19, 18, 311, 220, 386, 291, 327, 214, 145, 223, 137, 136, 119, 103, 404, 27, 32, 350, 400, 422, 423, 347, 73, 43, 55, 194, 189, 237, 230, 221, 247, 116, 242, 29, 37, 276, 345, 166, 179, 53, 87, 303, 90, 112, 245, 148, 41, 234, 208, 117, 193, 76, 236, 266, 130, 235, 251, 86, 31, 195, 144, 330, 63, 295, 70, 187, 79, 10, 33, 75, 361, 153, 284, 102, 21, 64, 45, 36, 181, 351, 272, 210, 243, 20, 98, 34, 16, 218. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Distributor
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Guinea, Honduras, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Mauritius, Mexico, Mongolia, Morocco, Nepal, Netherlands, New Zealand, Norway, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, T®rkiye, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.
Lot
Model Numbers: 722058; UDI-DI: 00884838059115; Serial Numbers: 337, 178, 249, 22, 202, 279, 254, 285, 71, 186, 82, 85, 162, 138, 24, 365, 203, 167, 146, 168, 304, 172, 200, 123, 129, 149, 259, 4, 197, 134, 38, 290, 171, 56, 264, 238, 199, 61, 156, 99, 65, 59, 109, 100, 305, 319, 215, 278, 124, 196, 405, 325, 265, 378, 277, 274, 94, 182, 366, 340, 190, 219, 373, 239, 23, 205, 257, 212, 111, 110, 399, 72, 332, 92, 142, 67, 106, 335, 360, 415, 11, 114, 151, 93, 232, 395, 292, 380, 183, 349, 261, 15, 375, 161, 78, 368, 9, 372, 288, 252, 320, 213, 293, 389, 169, 97, 336, 25, 143, 133, 192, 49, 412, 201, 89, 66, 163, 170, 77, 273, 35, 222, 229, 248, 334, 384, 40, 268, 240, 258, 289, 269, 226, 244, 48, 141, 391, 241, 164, 381, 281, 154, 246, 127, 315, 256, 233, 348, 204, 396, 7, 185, 95, 131, 377, 60, 68, 206, 101, 150, 333, 84, 398, 313, 3, 39, 344, 376, 382, 370, 387, 419, 394, 407, 413, 392, 411, 358, 409, 385, 416, 367, 379, 331, 418, 356, 421, 369, 420, 397, 357, 408, 393, 388, 108, 211, 126, 125, 301, 12, 30, 355, 354, 159, 139, 128, 81, 58, 96, 343, 177, 115, 19, 18, 311, 220, 386, 291, 327, 214, 145, 223, 137, 136, 119, 103, 404, 27, 32, 350, 400, 422, 423, 347, 73, 43, 55, 194, 189, 237, 230, 221, 247, 116, 242, 29, 37, 276, 345, 166, 179, 53, 87, 303, 90, 112, 245, 148, 41, 234, 208, 117, 193, 76, 236, 266, 130, 235, 251, 86, 31, 195, 144, 330, 63, 295, 70, 187, 79, 10, 33, 75, 361, 153, 284, 102, 21, 64, 45, 36, 181, 351, 272, 210, 243, 20, 98, 34, 16, 218;
Risk classification
Class II
Area (FDA)
device
Country
Netherlands

Related recalls

Questions about this recall

Was Allura Xper FD20/15; Model Numbers: 722058; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563; recalled?

Yes. FDA published recall Z-0010-2026 on October 8, 2025. Reported hazard: The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.

What should I do if I have this product?

Voluntary: Firm initiated · Letter