PRISM MEDICAL A-625, MAX. LOAD 625 lbs. 283 Kgs., Patient lift, non-A/C powered

Published October 18, 2017 · Recall initiated: September 8, 2017

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

During internal cycle testing of A-Series lifts Handicare has found that the A-Series lift strap for lifts manufactured between April 25, 2017 and June 12, 2017 may wear prematurely at max load (625 lbs). We have identified the root cause as a supplied component that was not to specification. If the strap wears during use there is a potential risk for the person being supported by the lift to drop.

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
Handicare Usa Inc
Distributor
Us and Canada
Lot
A6M0001526, A6M0001527, A6M0001528, A6M0001529, A6M0001530, A6M0001485, A6M0001486, A6M0001498
Risk classification
Class II
Area (FDA)
device
Country
United States