PRISM MEDICAL A-625, MAX. LOAD 625 lbs. 283 Kgs., Patient lift, non-A/C powered
Published October 18, 2017 · Recall initiated: September 8, 2017
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
During internal cycle testing of A-Series lifts Handicare has found that the A-Series lift strap for lifts manufactured between April 25, 2017 and June 12, 2017 may wear prematurely at max load (625 lbs). We have identified the root cause as a supplied component that was not to specification. If the strap wears during use there is a potential risk for the person being supported by the lift to drop.
What to do
Voluntary: Firm initiated · E-Mail
Recall details
- Manufacturer
- Handicare Usa Inc
- Distributor
- Us and Canada
- Lot
- A6M0001526, A6M0001527, A6M0001528, A6M0001529, A6M0001530, A6M0001485, A6M0001486, A6M0001498
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States