PRISM MEDICAL A-625, MAX. LOAD 625 lbs. 283 Kgs., Patient lift, non-A/C powered
Published October 18, 2017 · Recall initiated: September 8, 2017
Summary
FDA records medical device Class II recall Z-0013-2018, published October 18, 2017 and initiated September 8, 2017. Recalling firm: Handicare Usa Inc; Status: Terminated; Reported quantity: 8 units; Lot: A6M0001526, A6M0001527, A6M0001528, A6M0001529, A6M0001530, A6M0001485, A6M0001486, A6M0001498. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
During internal cycle testing of A-Series lifts Handicare has found that the A-Series lift strap for lifts manufactured between April 25, 2017 and June 12, 2017 may wear prematurely at max load (625 lbs). We have identified the root cause as a supplied component that was not to specification. If the strap wears during use there is a potential risk for the person being supported by the lift to drop.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Handicare Usa Inc
- Distributor
- Us and Canada
- Lot
- A6M0001526, A6M0001527, A6M0001528, A6M0001529, A6M0001530, A6M0001485, A6M0001486, A6M0001498
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABisoprolol Fumarate tablets, 5 mg, 30-count bottles, Rx only, Manufactured by: UNICHEM LABORATORIES, LTD.October 25, 2017
- FDASymbio Muc Eye, 0.17 fl.oz. 5mL bottles, Manufactured for and distributed in the US b: Biological Health GroupOctober 25, 2017
- FDAARCHITECT c4000 Cuvette Segment, List number 02P75-01. The firm name on the label is Abbott Laboratories,October 18, 2017
- FDACarina Mobile Lifts, Models: a) Carina350EM, REF 6060011 b) Carina350EML, REF 60600012 c) Carina350EE, REFOctober 18, 2017
- FDAWawa Fruit & Cheese, 6oz. (170g), 1 each, rigid plastic, clamshell container, sold per unit, shelf-life= 5October 18, 2017
- FDABirds Eye Baby Sweet Peas, Net Wt. 13 oz. (369g ), Keep Frozen. UPC 14500-02253.October 18, 2017
Questions about this recall
Was PRISM MEDICAL A-625, MAX. LOAD 625 lbs. 283 Kgs., Patient lift, non-A/C powered recalled?
Yes. FDA published recall Z-0013-2018 on October 18, 2017. Reported hazard: During internal cycle testing of A-Series lifts Handicare has found that the A-Series lift strap for lifts manufactured between April 25, 2017 and June 12, 2017 may wear prematurely at max load (625 lbs). We have identified the root cause as a supplied component that was not to specification. If the strap wears during use there is a potential risk for the person being supported by the lift to drop.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail