Carina Mobile Lifts, Models: a) Carina350EM, REF 6060011 b) Carina350EML, REF 60600012 c) Carina350EE, REF 60600009

Published October 18, 2017 · Recall initiated: July 28, 2016

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A limited amount of Carina mobile lifts have been delivered with a too small locking pin. If a locking pin smaller than the correct size is mounted on the Carina mobile lift, the locking pin is likely to cause increased wear on the plastic bushings. In case of continued use, these components may cause wear on the metal of the piston rod eye.

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
Handicare AB
Distributor
OH, PA, MD
Lot
a) Carina350EM, REF 6060011, Serial numbers: 661102000 - 661102160 b) Carina350EML, REF 60600012, Serial numbers: 661200500 - 661200584 c) Carina350EE, REF 60600009, Serial numbers: 660900500 -660900571
Risk classification
Class II
Area (FDA)
device
Country
Sweden