Beckman Coulter Dxl 9000 Access Immunoassay Analyzer, REF #C11137. An in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids.
Published October 16, 2024 · Recall initiated: April 18, 2024
Summary
FDA records medical device Class II recall Z-0025-2025, published October 16, 2024 and initiated April 18, 2024. Recalling firm: Beckman Coulter, Inc.; Status: Ongoing; Reported quantity: 164 units; Lot: UDI-DI: 15099590230623; U.S. serial numbers: 300284, 300286, 300294, 300266, 300291, 300223, 300260, 300285, 300292, 300195, 300227, 300228, 300205, 300207, 300206, 300182, 300192, and 300234. OUS serial numbers have been requested. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
When the DxI 9000 Access Immunoassay Analyzer with system software version 1.16.2 and prior is configured to automatically request a calibration order, the instrument can process the order with expired calibrator material. If the calibration curve is generated with an expired calibration material lot, this may lead to a hazardous situation of erroneously high or low patient results reported to the physicians.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Beckman Coulter, Inc.
- Distributor
- Distribution was made to FL, GA, IN, KS, NH, NJ, OK, PA, and TX. There was government distribution but no military distribution. Foreign distribution was made to Australia, Austria, Belgium, Brazil, Croatia, Czechia, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Province of China, and United Kingdom of Great Britain and Northern Ireland.
- Lot
- UDI-DI: 15099590230623; U.S. serial numbers: 300284, 300286, 300294, 300266, 300291, 300223, 300260, 300285, 300292, 300195, 300227, 300228, 300205, 300207, 300206, 300182, 300192, and 300234. OUS serial numbers have been requested.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAROTOR ASSY-SW40, Reference Number 331302, when the rotors are installed in the following centrifuges: OptimaOctober 28, 2026
- FDAROTOR ASSY-SW41, Reference Number 331362, when the rotors are installed in the following centrifuges: OptimaOctober 28, 2026
- FDAIQ200 SPRINT, REF: 700-3322,700-3322RFB, 700-3325, 700-3326, B48997; IQ200 SELECT, REF: 700-3340, 700-3345,October 21, 2026
- FDADxU 840m Iris, REF: C76947; DxU 850m Iris, REF: C49513October 21, 2026
- FDAiCast Covered Stent, 6MMx16MMx120CM, Model Number 42616October 16, 2024
- FDABD Synapsys Informatics Solution with software versions: 4.20, 4.30, 5.10, 5.20 and 5.30- IVD standaloneOctober 16, 2024
Questions about this recall
Was Beckman Coulter Dxl 9000 Access Immunoassay Analyzer, REF #C11137. An in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids. recalled?
Yes. FDA published recall Z-0025-2025 on October 16, 2024. Reported hazard: When the DxI 9000 Access Immunoassay Analyzer with system software version 1.16.2 and prior is configured to automatically request a calibration order, the instrument can process the order with expired calibrator material. If the calibration curve is generated with an expired calibration material lot, this may lead to a hazardous situation of erroneously high or low patient results reported to the physicians.
What should I do if I have this product?
Voluntary: Firm initiated · Letter