Beckman Coulter Dxl 9000 Access Immunoassay Analyzer, REF #C11137. An in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids.

Published October 16, 2024 · Recall initiated: April 18, 2024

Summary

FDA records medical device Class II recall Z-0025-2025, published October 16, 2024 and initiated April 18, 2024. Recalling firm: Beckman Coulter, Inc.; Status: Ongoing; Reported quantity: 164 units; Lot: UDI-DI: 15099590230623; U.S. serial numbers: 300284, 300286, 300294, 300266, 300291, 300223, 300260, 300285, 300292, 300195, 300227, 300228, 300205, 300207, 300206, 300182, 300192, and 300234. OUS serial numbers have been requested. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

When the DxI 9000 Access Immunoassay Analyzer with system software version 1.16.2 and prior is configured to automatically request a calibration order, the instrument can process the order with expired calibrator material. If the calibration curve is generated with an expired calibration material lot, this may lead to a hazardous situation of erroneously high or low patient results reported to the physicians.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Beckman Coulter, Inc.
Distributor
Distribution was made to FL, GA, IN, KS, NH, NJ, OK, PA, and TX. There was government distribution but no military distribution. Foreign distribution was made to Australia, Austria, Belgium, Brazil, Croatia, Czechia, Egypt, France, Germany, Hungary, Ireland, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Province of China, and United Kingdom of Great Britain and Northern Ireland.
Lot
UDI-DI: 15099590230623; U.S. serial numbers: 300284, 300286, 300294, 300266, 300291, 300223, 300260, 300285, 300292, 300195, 300227, 300228, 300205, 300207, 300206, 300182, 300192, and 300234. OUS serial numbers have been requested.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Beckman Coulter Dxl 9000 Access Immunoassay Analyzer, REF #C11137. An in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids. recalled?

Yes. FDA published recall Z-0025-2025 on October 16, 2024. Reported hazard: When the DxI 9000 Access Immunoassay Analyzer with system software version 1.16.2 and prior is configured to automatically request a calibration order, the instrument can process the order with expired calibrator material. If the calibration curve is generated with an expired calibration material lot, this may lead to a hazardous situation of erroneously high or low patient results reported to the physicians.

What should I do if I have this product?

Voluntary: Firm initiated · Letter