ROTOR ASSY-SW41, Reference Number 331362, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima XPN-90 (Model Numbers A99842 and A94468), Optima XPN-80 (Model Numbers A99839 and A95765), Optima XE-100 (Model Numbers A99836 and A94516), and Optima XE-90 (Model Numbers A99833 and A94471)
Published October 28, 2026 · Recall initiated: September 21, 2026
Summary
FDA records medical device Class II recall Z-0202-2027, published October 28, 2026 and initiated September 21, 2026. Recalling firm: Beckman Coulter, Inc.; Status: Ongoing; Reported quantity: 336 units; Lot: Serial Numbers: 24U15250 to 26U15766; Optima UDI-DI: A99846 - 15099590416324, A99842 - 15099590415754, A99839 - 15099590415822, A99836 - 15099590415846, A99833 - 15099590415860. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Rotors have been found with a non-conforming microstructure, which may result in some rotors breaking into two or more pieces during use. Rotors that break into multiple pieces during use may result in significant damage to materials (i.e., samples) being processed in the centrifuge and to the centrifuge instrument.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Beckman Coulter, Inc.
- Distributor
- US Nationwide distribution.
- Lot
- Serial Numbers: 24U15250 to 26U15766; Optima UDI-DI: A99846 - 15099590416324, A99842 - 15099590415754, A99839 - 15099590415822, A99836 - 15099590415846, A99833 - 15099590415860
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAROTOR ASSY-SW40, Reference Number 331302, when the rotors are installed in the following centrifuges: OptimaOctober 28, 2026
- FDACareLink Clinic, REF: MMT-7350October 28, 2026
- FDABrand Name: VACUETTE Product Name: LH Lithium Heparin Separator Model/Catalog Number: 454008P ProductOctober 28, 2026
- FDAIQ200 SPRINT, REF: 700-3322,700-3322RFB, 700-3325, 700-3326, B48997; IQ200 SELECT, REF: 700-3340, 700-3345,October 21, 2026
- FDADxU 840m Iris, REF: C76947; DxU 850m Iris, REF: C49513October 21, 2026
- FDACOULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent inFebruary 11, 2026
Questions about this recall
Was ROTOR ASSY-SW41, Reference Number 331362, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima XPN-90 (Model Numbers A99842 and A94468), Optima XPN-80 (Model Numbers A99839 and A95765), Optima XE-100 (Model Numbers A99836 and A94516), and Optima XE-90 (Model Numbers A99833 and A94471) recalled?
Yes. FDA published recall Z-0202-2027 on October 28, 2026. Reported hazard: Rotors have been found with a non-conforming microstructure, which may result in some rotors breaking into two or more pieces during use. Rotors that break into multiple pieces during use may result in significant damage to materials (i.e., samples) being processed in the centrifuge and to the centrifuge instrument.
What should I do if I have this product?
Voluntary: Firm initiated · Letter