PREMIER(R) Anterior Cervical Plate System SCREW BLOCK, REF 1860503, QTY: 1 EA, Rx only, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee 38132. Orthopedic. This instrument is a precision device which may incorporate a measuring function and has uses as described on the label. Unless labeled for single use, this instrument may be re-used.
Published October 17, 2012 · Recall initiated: August 30, 2012
Summary
FDA records medical device Class II recall Z-0038-2013, published October 17, 2012 and initiated August 30, 2012. Recalling firm: Medtronic Sofamor Danek USA Inc; Status: Terminated; Reported quantity: 247 units; Lot: All lots except those starting with SY. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Lines on the screw gauge may not placed on the caddy properly which could result in improper measurement of screws. An incorrect gauge could measure a screw to be approximately 1 mm shorter than its actual length.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Medtronic Sofamor Danek USA Inc
- Distributor
- Worldwide Distribution-USA (nationwide) and the countries of India, Netherlands, Hong Kong, S. L Korea, Singapore, Canada, Australia, Japan, and Mexico.
- Lot
- All lots except those starting with SY
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAGrafton" DBM and Grafton Plus" DBM contain demineralized human bone tissue combined with an inert additive toOctober 21, 2026
- FDAGrafton" DBM and Grafton Plus" DBM contain demineralized human bone tissue combined with an inert additive toOctober 21, 2026
- FDAGrafton DMB Matrix, Model Numbers: a) T42200, b) T42210; Demineralized Bone MatrixNovember 1, 2023
- FDAEndoskeleton" TCS Interbody System with Titan nanoLOCK" Surface Technology, NanoLock, 6 Deg, Interbody,October 20, 2021
- FDAATLANTIS(R) Anterior Cervical Plate System FIXED ANGLE SCREW MODULE, REF 176-504, QTY: 1 EA, Rx only,October 17, 2012
- FDAIDS Urine BETA CrossLaps ELISA, Catalog Number AC-05F1. Type of packaging: Cardboard Box containing plasticOctober 17, 2012
Questions about this recall
Was PREMIER(R) Anterior Cervical Plate System SCREW BLOCK, REF 1860503, QTY: 1 EA, Rx only, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee 38132. Orthopedic. This instrument is a precision device which may incorporate a measuring function and has uses as described on the label. Unless labeled for single use, this instrument may be re-used. recalled?
Yes. FDA published recall Z-0038-2013 on October 17, 2012. Reported hazard: Lines on the screw gauge may not placed on the caddy properly which could result in improper measurement of screws. An incorrect gauge could measure a screw to be approximately 1 mm shorter than its actual length.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit