STAR Excimer Laser System, Model: STAR S2, Catalog Number: 0030-1479, Product is manufactured and distributed by AMO Manufacturing USA, LLC, Milpitas, CA The STAR Excimer Laser Systems are indicated for laser assisted in situ keratomileusis (LASIK), photorefractive keratectomy (PRK), and phototherapeutic keratectomy (PTK) eye surgery in certain patient populations.
Published October 30, 2013 · Recall initiated: September 16, 2013
Summary
FDA records medical device Class II recall Z-0046-2014, published October 30, 2013 and initiated September 16, 2013. Recalling firm: Abbott Medical Optics, Inc.; Status: Terminated; Reported quantity: 834 units in total; Lot: All units of these models: STAR S2 Excimer Laser System; STAR S3 ActiveTrak; STAR S4 with Variable Spot Scanning; and STAR S41R; There are no expiration dates for the STAR Excimer Laser Systems. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
This corrective action will address an identified failure of the lamp and illumination system within the laser unit caused by an unexpected failure of a thyristor component on the AC Power Distribution Printed Circuit Board (PCB).
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Abbott Medical Optics, Inc.
- Distributor
- Worldwide distribution: US (nationwide) including Puerto Rico and countries of: Asia Pacific, Canada, Europe, Africa, Middle East, Japan, and Latin America.
- Lot
- All units of these models: STAR S2 Excimer Laser System; STAR S3 ActiveTrak; STAR S4 with Variable Spot Scanning; and STAR S41R; There are no expiration dates for the STAR Excimer Laser Systems.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOptiMedica Catalys Precision Laser System; Model No(s). Catalys-U (US), and Catalys-I (Outside US). ProductApril 29, 2015
- FDASTAR Excimer Laser System, Model STAR S4 IR, Catalog Number: 0030-4864, Product is manufactured andOctober 30, 2013
- FDASTAR ActiveTrak Excimer Laser System, Model: STAR S3, Catalog Number: 0030-2450, Product is manufactured andOctober 30, 2013
- FDASTAR Excimer Laser System with Variable Spot Scanning, Model: STAR S4, Catalog Number: 0030-4077, Product isOctober 30, 2013
- FDAMerit Medical Performa Angiographic Catheter 5F Multipack. Part numbers 7560-43 and 7560-G3.October 30, 2013
- FDASTERRAD 100NX Cassettes The STERRAD Sterilization System is a low temperature general purpose sterilizer usedOctober 30, 2013
Questions about this recall
Was STAR Excimer Laser System, Model: STAR S2, Catalog Number: 0030-1479, Product is manufactured and distributed by AMO Manufacturing USA, LLC, Milpitas, CA The STAR Excimer Laser Systems are indicated for laser assisted in situ keratomileusis (LASIK), photorefractive keratectomy (PRK), and phototherapeutic keratectomy (PTK) eye surgery in certain patient populations. recalled?
Yes. FDA published recall Z-0046-2014 on October 30, 2013. Reported hazard: This corrective action will address an identified failure of the lamp and illumination system within the laser unit caused by an unexpected failure of a thyristor component on the AC Power Distribution Printed Circuit Board (PCB).
What should I do if I have this product?
Voluntary: Firm initiated · Letter