Alere Triage® TOX Drug Screen Control 1 Product Generic Name: Quality Control Material Description of the product: Good Laboratory Practice suggests that external controls should be tested with each new lot or shipment of test materials, or every 30 days, and as otherwise required by a laboratorys standard quality control procedures. The Alere Triage® Drug Screen Controls, when used in accordance with the test procedures described in the Alere Triage® TOX Drug Screen package insert, provide this capability. The results obtained for the controls are to be compared with the assigned values given on the Expected Values Card, accompanying the package insert, to determine if the procedure is within control limits. Alere Triage® TOX Drug Screen Control 1 or Alere Triage® TOX Drug Screen Control 2 consist of human urine matrix containing parent compounds and/or major metabolites/analogues of drugs of abuse. 510(k)/IDE/PMA number: K060788 Type of Packaging: 5 vials per level in a labeled kit box. Indications for use: The Alere Triage® TOX Drug Screen Controls are to be used with the Alere Triage® TOX Drug Screen tests and Alere Triage® Meters to assist the laboratory in monitoring test performance.
Published October 30, 2013 · Recall initiated: September 23, 2013
Summary
FDA records medical device Class III recall Z-0053-2014, published October 30, 2013 and initiated September 23, 2013. Recalling firm: Alere San Diego, Inc.; Status: Terminated; Reported quantity: 38; Lot: Part Number 94413. Lot Number C2681A. Class III means the product is not likely to cause adverse health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Alere San Diego, Inc. initiated a voluntary recall of Alere Triage TOX Drug Screen Control 1 (PN 94413, Lot C2681A) because the CODE CHIP that was provided with this lot causes the control to fail when run in the QC Sample Mode using Alere Triage TOX Drug Screeen products. The Alere Triage TOX Drug Screen test devices and patient results are not affected.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Alere San Diego, Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) and countries of: South Africa and Canada.
- Lot
- Part Number 94413. Lot Number C2681A.
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMcKesson CONSULT hCG Urine Cassette, Model/ Part: FHC-A102/5001; Alere hCG Cassette (20 mIU/mL) in Urine,October 7, 2026
- FDAFisher Scientific Sure-Vue Urine hCG Strip, Model/ Part: FHC-101/23900527; McKesson CONSULT hCG Urine TestOctober 7, 2026
- FDAFisher Scientific Sure-Vue Urine hCG, Model/ Part: FHC-102/23900526; Henry Schein, Inc. One Step + UrineOctober 7, 2026
- FDAFisher Scientific Sure-Vue Serum/Urine hCG, Model/ Part: FHC-202/23900529, FHC-202/23900528; Beckman CoulterOctober 7, 2026
- FDAAcuson SC2000 volume imaging ultrasound systems at software version 3.0. Mfg name: Siemens Medical SolutionsOctober 30, 2013
- FDAIntegra Licox Brain Tissue Oxygen Monitoring; 1 x Complete Brain Probe Kit REF IP1.P Rx Only, ManufacturerOctober 30, 2013
Questions about this recall
Was Alere Triage® TOX Drug Screen Control 1 Product Generic Name: Quality Control Material Description of the product: Good Laboratory Practice suggests that external controls should be tested with each new lot or shipment of test materials, or every 30 days, and as otherwise required by a laboratorys standard quality control procedures. The Alere Triage® Drug Screen Controls, when used in accordance with the test procedures described in the Alere Triage® TOX Drug Screen package insert, provide this capability. The results obtained for the controls are to be compared with the assigned values given on the Expected Values Card, accompanying the package insert, to determine if the procedure is within control limits. Alere Triage® TOX Drug Screen Control 1 or Alere Triage® TOX Drug Screen Control 2 consist of human urine matrix containing parent compounds and/or major metabolites/analogues of drugs of abuse. 510(k)/IDE/PMA number: K060788 Type of Packaging: 5 vials per level in a labeled kit box. Indications for use: The Alere Triage® TOX Drug Screen Controls are to be used with the Alere Triage® TOX Drug Screen tests and Alere Triage® Meters to assist the laboratory in monitoring test performance. recalled?
Yes. FDA published recall Z-0053-2014 on October 30, 2013. Reported hazard: Alere San Diego, Inc. initiated a voluntary recall of Alere Triage TOX Drug Screen Control 1 (PN 94413, Lot C2681A) because the CODE CHIP that was provided with this lot causes the control to fail when run in the QC Sample Mode using Alere Triage TOX Drug Screeen products. The Alere Triage TOX Drug Screen test devices and patient results are not affected.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit