Back to search
FDAFDA · Food and drugs·Recall number Z-0054-2014

Integra Licox Brain Tissue Oxygen Monitoring; 1 x Complete Brain Probe Kit REF IP1.P Rx Only, Manufacturer GMS- Gesellschaft fur medizinische, Sondentechnik mbH Dofstrasse 2 24247 Melkendorf, Germany Distributed by Integra Neurosciences 311 Enterprise Drive Plainsboro, NJ 08536 USA. Used in direct monitoring of the partial pressure of oxygen.

Published October 30, 2013 · Recall initiated: June 11, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Integra LifeSciences Corporation has determined that there is a possibility that some lots of Licox CMP Brain Tissue Oxygen and Temperature Catheter Kits contain a hex wrench that does not loosen the set screw on the drill bit.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Integra LifeSciences Corp.
Distributor
Worldwide distribution: US (nationwide) and countries of: AR, BR, CH, CL,CR. DE, ES, FI, FR, GB, IL, IT, MX, PA, PL, PT, RU, TW, and ZA.
Lot
K040235 Lot 160812
Risk classification
Class II
Area (FDA)
device
Country
United States