FDAFDA · Food and drugs·Recall number Z-0071-2027

VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB007

Published October 14, 2026 · Recall initiated: August 31, 2026

Summary

FDA records medical device Class II recall Z-0071-2027, published October 14, 2026 and initiated August 31, 2026. Recalling firm: Steris Corporation; Status: Ongoing; Reported quantity: 409; Lot: UDI: 10724995220331 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Steris Corporation
Distributor
Worldwide - US Nationwide distribution.
Lot
UDI: 10724995220331 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB007 recalled?

Yes. FDA published recall Z-0071-2027 on October 14, 2026. Reported hazard: Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction

What should I do if I have this product?

Voluntary: Firm initiated · Letter