RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020117
Published October 21, 2026 · Recall initiated: September 21, 2026
Summary
FDA records medical device Class II recall Z-0156-2027, published October 21, 2026 and initiated September 21, 2026. Recalling firm: Steris Corporation; Status: Ongoing; Reported quantity: 106758 units; Lot: UDI: 00724995245962 Lot numbers with expiration dates: 688019 10/19/2027 688021 10/19/2027 688023 10/19/2027 688025 10/19/2027 688576 10/26/2027 688580 10/26/2027 688582 10/26/2027 688584 10/26/2027 688586 10/26/2027 688588 10/26/2027 689173 11/2/2027 689175 11/2/2027 689177 11/2/2027 689179 11/2/2027 689983 11/10/2027 689985 11/10/2027 689987 11/10/2027 690299 11/16/2027 690301 11/16/2027 690995 11/24/2027 690997 11/24/2027 690999 11/24/2027 691001 11/24/2027 691741 12/1/2027 691743 12/1/2027 691745 12/8/2027 692203 12/8/2027 692206 12/8/2027 692208 12/8/2027 692210 12/8/2027 692213 12/8/2027 692215 12/8/2027 692217 12/9/2027 692572 12/10/2027 692574 12/10/2027 692576 12/10/2027 692578 12/10/2027 692580 12/10/2027 692582 12/10/2027 692730 12/14/2027 692732 12/14/2027 692734 12/14/2027 692736 12/14/2027 692738 12/14/2027 692740 12/14/2027 692742 12/14/2027 692744 12/14/2027 694359 12/30/2027 694361 12/30/2027 694363 12/30/2027. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Steris Corporation
- Distributor
- Worldwide - US Nationwide distribution.
- Lot
- UDI: 00724995245962 Lot numbers with expiration dates: 688019 10/19/2027 688021 10/19/2027 688023 10/19/2027 688025 10/19/2027 688576 10/26/2027 688580 10/26/2027 688582 10/26/2027 688584 10/26/2027 688586 10/26/2027 688588 10/26/2027 689173 11/2/2027 689175 11/2/2027 689177 11/2/2027 689179 11/2/2027 689983 11/10/2027 689985 11/10/2027 689987 11/10/2027 690299 11/16/2027 690301 11/16/2027 690995 11/24/2027 690997 11/24/2027 690999 11/24/2027 691001 11/24/2027 691741 12/1/2027 691743 12/1/2027 691745 12/8/2027 692203 12/8/2027 692206 12/8/2027 692208 12/8/2027 692210 12/8/2027 692213 12/8/2027 692215 12/8/2027 692217 12/9/2027 692572 12/10/2027 692574 12/10/2027 692576 12/10/2027 692578 12/10/2027 692580 12/10/2027 692582 12/10/2027 692730 12/14/2027 692732 12/14/2027 692734 12/14/2027 692736 12/14/2027 692738 12/14/2027 692740 12/14/2027 692742 12/14/2027 692744 12/14/2027 694359 12/30/2027 694361 12/30/2027 694363 12/30/2027
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020116October 21, 2026
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020115October 21, 2026
- FDAREF: P65 ST862 CoLag SCREW, FT, 6.7X62.5 mmOctober 21, 2026
- FDAGeneric Name: GORE VIABIL Short Wire Biliary Endoprosthesis; Model/Catalog Number: VSWVH0808 and VSWVN1006October 21, 2026
- FDAVERIFY Biological Steam Test Pack Model/Catalog Number: S3069October 14, 2026
- FDAVERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB007October 14, 2026
Questions about this recall
Was RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020117 recalled?
Yes. FDA published recall Z-0156-2027 on October 21, 2026. Reported hazard: If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.
What should I do if I have this product?
Voluntary: Firm initiated · Letter