VERIFY Biological Steam Test Pack Model/Catalog Number: S3065
Published October 14, 2026 · Recall initiated: August 31, 2026
Summary
FDA records medical device Class II recall Z-0074-2027, published October 14, 2026 and initiated August 31, 2026. Recalling firm: Steris Corporation; Status: Ongoing; Reported quantity: 217; Lot: UDI: 10724995220300 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Steris Corporation
- Distributor
- Worldwide - US Nationwide distribution.
- Lot
- UDI: 10724995220300 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020117October 21, 2026
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020115October 21, 2026
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020116October 21, 2026
- FDAVERIFY Biological Steam Test Pack Model/Catalog Number: S3069October 14, 2026
- FDAVERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3061October 14, 2026
- FDAVERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3060October 14, 2026
Questions about this recall
Was VERIFY Biological Steam Test Pack Model/Catalog Number: S3065 recalled?
Yes. FDA published recall Z-0074-2027 on October 14, 2026. Reported hazard: Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
What should I do if I have this product?
Voluntary: Firm initiated · Letter