FDAFDA · Food and drugs·Recall number Z-0089-2027

Product Name: Nihon Kohden CNS-6200 Series Central Nurse Station, Model/Catalog Number: CN-6201 Software Version: software versions: 01-03, 01-04, 01-05, 01-06, 02-10, 02-11, and 02-40 Product Description: The device is intended for use by medical professionals to provide cardiac and vital signs monitoring for multiple patients within a medical facility. The CNS-6200 Series Central Nurse Station will display and record physiological data from up to forty telemetry receiver/transmitters and generates an alarm when a measured parameter falls outside a pre-set limit or when life threatening arrhythmia is detected. Arrhythmia detection and alarm determination are functions of the telemetry receivers/transmitters or individual bedside monitor. Component: No

Published October 14, 2026 · Recall initiated: September 8, 2026

Summary

FDA records medical device Class II recall Z-0089-2027, published October 14, 2026 and initiated September 8, 2026. Recalling firm: Nihon Kohden America LLC; Status: Ongoing; Reported quantity: 808; Lot: Lot Code: CNS-6201 (PU-621RA), SW versions: 01-03, 01-04, 01-05, 01-06, 02-10, 02-11, and 02-40. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Due to a cybersecurity issue associate with a NULL Pointer Dereferences vulnerabiity. If exploited would potentially cause disruption of the centralized monitoring of multiple beds and the visualization of alarms.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Nihon Kohden America LLC
Distributor
US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN. KS, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NL, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, VT, QA, WA, WI, WV and WY.
Lot
Lot Code: CNS-6201 (PU-621RA), SW versions: 01-03, 01-04, 01-05, 01-06, 02-10, 02-11, and 02-40
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Product Name: Nihon Kohden CNS-6200 Series Central Nurse Station, Model/Catalog Number: CN-6201 Software Version: software versions: 01-03, 01-04, 01-05, 01-06, 02-10, 02-11, and 02-40 Product Description: The device is intended for use by medical professionals to provide cardiac and vital signs monitoring for multiple patients within a medical facility. The CNS-6200 Series Central Nurse Station will display and record physiological data from up to forty telemetry receiver/transmitters and generates an alarm when a measured parameter falls outside a pre-set limit or when life threatening arrhythmia is detected. Arrhythmia detection and alarm determination are functions of the telemetry receivers/transmitters or individual bedside monitor. Component: No recalled?

Yes. FDA published recall Z-0089-2027 on October 14, 2026. Reported hazard: Due to a cybersecurity issue associate with a NULL Pointer Dereferences vulnerabiity. If exploited would potentially cause disruption of the centralized monitoring of multiple beds and the visualization of alarms.

What should I do if I have this product?

Voluntary: Firm initiated · Letter