Codman EDS 3 Drainage System with Ventricular Catheter (External Drainage System) Catalog Number: 82-1730 Indicated for draining cerebrospinal fluid (CFS) and other fluids of similar physical characteristics as a means of reducing intracranial pressure and CSF volume when the insertion of a permanent, internal shunt is not indicated
Published October 29, 2014 · Recall initiated: September 15, 2014
Summary
FDA records medical device Class II recall Z-0089-2015, published October 29, 2014 and initiated September 15, 2014. Recalling firm: Codman & Shurtleff, Inc.; Status: Terminated; Reported quantity: 66,120 units; Lot: All lots with expiration on or before August 2017 (2017-08). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
tubing within the system that drains CSF may leak or disconnect from the joints and may result in over- or under-drainage of CSF from the ventricular system or introduction of air into the ventricular system (pneumocephalus)
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Codman & Shurtleff, Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) and Internationally to: United Arab Emirates Argentina Austria Australia Aruba Belgium Bahrain Bolivia Brazil Bahamas Canada Switzerland Chile People's Republic of China Columbia Costa Rica Cyprus Czech Republic Germany Algeria Estonia Egypt Spain Finland France Gabon Great Britain French Guiana Greece Guatemala Hong Kong Croatia Hungary Ireland Israel India Iran Italy Jordan Kenya Kuwait Kazakhstan Lebanon Lithuania Luxembourg Latvia Mauritius Mexico Malaysia Mozambique New Caledonia Nigeria Netherlands Norway New Zealand Oman Panama Peru Poland Palestine Portugal Reunion Romania Serbia Russian Federation Rwanda Saudi Arabia Sweden Singapore Slovenia Slovakia Tunisia Turkey Trinadad And Tobago Taiwan Ukraine Uruguay Venezuela Mayotte South Africa
- Lot
- All lots with expiration on or before August 2017 (2017-08)
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACodman EDS 3 Drainage System (no Ventricular Catheter) (External Drainage System) Catalog Number: 82-1731October 29, 2014
- FDACodman Lumbar Drainage Catheter Kit II with EDS 3 (External Drainage System) Catalog Number: 82-1738 IndicatedOctober 29, 2014
- FDABoston Scientific TELIGEN, Implantable Cardiac Defibrillator models E102, E110, the following models are notOctober 29, 2014
- FDABoston Scientific COGNIS CRT-D, Cardiac Resynchronization Therapy Defibrillator, models N118, N119. TheOctober 29, 2014
- FDACodman Certas - In Line Valve only with SIPHONGUARD® Device; Product Code: 82-8804 The Codman CertasAugust 20, 2014
- FDACodman Certas - In Line Valve Only; Product Code: 82-8800 The Codman Certas Programmable valve is anAugust 20, 2014
Questions about this recall
Was Codman EDS 3 Drainage System with Ventricular Catheter (External Drainage System) Catalog Number: 82-1730 Indicated for draining cerebrospinal fluid (CFS) and other fluids of similar physical characteristics as a means of reducing intracranial pressure and CSF volume when the insertion of a permanent, internal shunt is not indicated recalled?
Yes. FDA published recall Z-0089-2015 on October 29, 2014. Reported hazard: tubing within the system that drains CSF may leak or disconnect from the joints and may result in over- or under-drainage of CSF from the ventricular system or introduction of air into the ventricular system (pneumocephalus)
What should I do if I have this product?
Voluntary: Firm initiated · Letter