Boston Scientific TELIGEN, Implantable Cardiac Defibrillator models E102, E110, the following models are not available in the US: models E103, E111, F102, F103, F110, F111. Sterilized using ethylene oxide. Product Usage: This family of implantable cardioverter defibrillators (ICDs) provide a variety of therapies including: -Ventricular tachyarrhythmia therapy, which is used to treat rhythms associated with sudden cardiac death such as VT and VF; -Bradycardia pacing, including adaptive rate pacing, to detect and treat bradyarrhythmias and to provide cardiac rate support after defibrillation therapy. Boston Scientific implantable cardioverter defibrillators (ICDs) are indicated to provide ventricular antitachycardia pacing (ATP) and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias

Published October 29, 2014 · Recall initiated: September 17, 2014

Summary

FDA records medical device Class II recall Z-0088-2015, published October 29, 2014 and initiated September 17, 2014. Recalling firm: Boston Scientific CRM Corp; Status: Terminated; Reported quantity: 31,144 devices total; Lot: Model E102 - Serial Numbers: 240307, 240308, 240310, 240312, 240313, 240318, 240319, 240320, 240321, 240324, 240325, 240326, 240327, 240329, 240330, 240332, 240336, 240338, 240340, 240342, 240344, 240345, 240346, 240348, 240349, 240350, 240351, 240354, 240355, 240356, 240358, 240359, 240360, 240363, 240365, 240366, 240367, 240368, 240369, 240371, 240372, 240373, 240374, 240375, 240376, 240378, 240380, 240383, 240384, 240386, 240389, 240391, 240393, 240397, 240398, 240399, 240401, 240403, 240404, 240405, 240406, 240407, 240408, 240410, 240414, 240415, 240416, 240418, 240421, 240422, 240423, 240424, 240426, 240432, 240433, 240434, 240438, 240439, 240441, 240443, 240444, 240448, 240450, 240451, 240452, 240454, 240455, 240456, 240457, 240459, 240460, 240461, 240462, 240466, 240468, 240471, 240472, 240473, 240474, 240475, 240478, 240479, 240480, 240484, 240485, 240488, 240489, 240490, 240491, 240492, 240493, 240496, 240497, 240500, 240501, 240503, 240505, 240508, 240510, 240511, 240512, 240517, 240519, 240521, 240523, 240524, 240525, 240527, 240529, 240530, 240531, 240533, 240535, 240536, 240537, 240540, 240542, 240544, 240547, 240548, 240549, 240550, 240551, 240552, 240555, 240556, 240557, 240558, 240559, 240561, 240562, 240563, 240566, 240573, 240575, 240576, 240579, 240582, 240583, 240584, 240586, 240587, 240589, 240590, 240591, 240592, 240593, 240594, 240596, 240597, 240598, 240599, 240600, 240602, 240605, 240606, 240607, 240608, 240609, 240610, 240611, 240612, 240615, 240616, 240618, 240619, 240622, 240623, 240624, 240627, 240628, 240630, 240633, 240635, 240636, 240637, 240639, 240640, 240641, 240642, 240644, 240645, 240646, 240647, 240648, 240649, 240650, 240652, 240653, 240655, 240656, 240657, 240658, 240659, 240660, 240663, 240664, 240665, 240666, 240668, 240669, 240672, 240674, 240678, 240679, 240680, 240681, 240687, 240688, 240690, 240692, 240699, 240701, 240703, 240705, 240707, 240708, 240709, 240712, 240713, 240715, 240717, 240718, 240719, 240720, 240722, 240723, 240725, 240726, 240727, 240728, 240729, 240730, 240731, 240732, 240736, 240737, 240738, 240740, 240741, 240743, 240744, 240747, 240749, 240750, 240751, 240752, 240755, 240757, 240758, 240759, 240761, 240762, 240763, 240764, 240765, 240766, 240767, 240768, 240769, 240770, 240771, 240775, 240777, 240778, 240779, 240780, 240781, 240782, 240788, 240791, 240792, 240794, 240797, 240798, 240799, 240800, 240801, 240802, 240803, 240804, 240805, 240806, 240807, 240808, 240810, 240814, 240818, 240821, 240822, 240826, 240828, 240829, 240830, 240832, 240833, 240834, 240835, 240836, 240837, 240840, 240841, 240844, 240845, 240848, 240849, 240850, 240851, 240852, 240853, 240854, 240855, 240858, 240859, 240861, 240862, 240863, 240864, 240865, 240868, 240869, 240870, 240872, 240873, 240874, 240875, 240876, 240877, 240878, 240879, 240881, 240882, 240883, 240884, 240886, 240889, 240890, 240891, 240893, 240894, 240895, 240896, 240897, 240898, 240899, 240900, 240901, 240902, 240903, 240905, 240906, 240910, 240912, 240913, 240914, 240916, 240917, 240918, 240924, 240928, 240932, 240934, 240939, 240941, 240942, 240943, 240944, 240945, 240946, 240949, 240950, 240952, 240953, 240954, 240955, 240956, 240957, 240959, 240960, 240961, 240962, 240965, 240966, 240968, 240969, 240971, 240973, 240975, 240976, 240977, 240979, 240980, 240981, 240982, 240985, 240988, 240990, 240991, 240992, 240993, 240995, 240996, 240997, 240999, 241000, 241001, 241003, 241004, 241005, 241006, 241007, 241010, 241012, 241014, 241017, 241020, 241022, 241023, 241026, 241027, 241028, 241029, 241030, 241031, 241032, 241033, 241034, 241037, 241038, 241039, 241041, 241042, 241044, 241045, 241048, 241050, 241051, 241053, 241054, 241056, 241058, 241060, 241061, 241063, 241064, 241065, 241069, 241074, 241076, 241077, 241079, 241082, 241083, 241085, 241086, 241088, 241089, 241090, 241092, 241093, 241095, 241096, 241097, 241098, 241100, 241101, 241102, 241103, 241104, 241106, 241107, 241108, 241109, 241112, 241113, 241114, . Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Boston Scientific identified a subset of COGNISTM CRT-Ds and TELIGENTM ICDs that had experienced an increased rate of premature battery depletion due to compromised performance of a low voltage (LV) capacitor. Boston Scientific updated their August 2013 Physician communication to include additional devices that may experience premature battery depletion due to compromised performance of a low volt

What to do

Voluntary: Firm initiated · Visit

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Boston Scientific CRM Corp
Distributor
Worldwide Distribution - US Nationwide including American Samoa, Guam, Puerto Rico, and U.S. - Virgin Islands. OUS countries to include: Argentina, Aruba, Australia, Austria, Bahamas, Barbados, Belgium, Bermuda, Brazil, Canada, Cayman Islands, Chile, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, Germany, Greece, Guadeloupe, Haiti, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Korea, Kuwait, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Monaco, Mozambique, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, North Yemen, Norway, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan R O C, Thailand, Trinidad and Tobago, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela.
Lot
Model E102 - Serial Numbers: 240307, 240308, 240310, 240312, 240313, 240318, 240319, 240320, 240321, 240324, 240325, 240326, 240327, 240329, 240330, 240332, 240336, 240338, 240340, 240342, 240344, 240345, 240346, 240348, 240349, 240350, 240351, 240354, 240355, 240356, 240358, 240359, 240360, 240363, 240365, 240366, 240367, 240368, 240369, 240371, 240372, 240373, 240374, 240375, 240376, 240378, 240380, 240383, 240384, 240386, 240389, 240391, 240393, 240397, 240398, 240399, 240401, 240403, 240404, 240405, 240406, 240407, 240408, 240410, 240414, 240415, 240416, 240418, 240421, 240422, 240423, 240424, 240426, 240432, 240433, 240434, 240438, 240439, 240441, 240443, 240444, 240448, 240450, 240451, 240452, 240454, 240455, 240456, 240457, 240459, 240460, 240461, 240462, 240466, 240468, 240471, 240472, 240473, 240474, 240475, 240478, 240479, 240480, 240484, 240485, 240488, 240489, 240490, 240491, 240492, 240493, 240496, 240497, 240500, 240501, 240503, 240505, 240508, 240510, 240511, 240512, 240517, 240519, 240521, 240523, 240524, 240525, 240527, 240529, 240530, 240531, 240533, 240535, 240536, 240537, 240540, 240542, 240544, 240547, 240548, 240549, 240550, 240551, 240552, 240555, 240556, 240557, 240558, 240559, 240561, 240562, 240563, 240566, 240573, 240575, 240576, 240579, 240582, 240583, 240584, 240586, 240587, 240589, 240590, 240591, 240592, 240593, 240594, 240596, 240597, 240598, 240599, 240600, 240602, 240605, 240606, 240607, 240608, 240609, 240610, 240611, 240612, 240615, 240616, 240618, 240619, 240622, 240623, 240624, 240627, 240628, 240630, 240633, 240635, 240636, 240637, 240639, 240640, 240641, 240642, 240644, 240645, 240646, 240647, 240648, 240649, 240650, 240652, 240653, 240655, 240656, 240657, 240658, 240659, 240660, 240663, 240664, 240665, 240666, 240668, 240669, 240672, 240674, 240678, 240679, 240680, 240681, 240687, 240688, 240690, 240692, 240699, 240701, 240703, 240705, 240707, 240708, 240709, 240712, 240713, 240715, 240717, 240718, 240719, 240720, 240722, 240723, 240725, 240726, 240727, 240728, 240729, 240730, 240731, 240732, 240736, 240737, 240738, 240740, 240741, 240743, 240744, 240747, 240749, 240750, 240751, 240752, 240755, 240757, 240758, 240759, 240761, 240762, 240763, 240764, 240765, 240766, 240767, 240768, 240769, 240770, 240771, 240775, 240777, 240778, 240779, 240780, 240781, 240782, 240788, 240791, 240792, 240794, 240797, 240798, 240799, 240800, 240801, 240802, 240803, 240804, 240805, 240806, 240807, 240808, 240810, 240814, 240818, 240821, 240822, 240826, 240828, 240829, 240830, 240832, 240833, 240834, 240835, 240836, 240837, 240840, 240841, 240844, 240845, 240848, 240849, 240850, 240851, 240852, 240853, 240854, 240855, 240858, 240859, 240861, 240862, 240863, 240864, 240865, 240868, 240869, 240870, 240872, 240873, 240874, 240875, 240876, 240877, 240878, 240879, 240881, 240882, 240883, 240884, 240886, 240889, 240890, 240891, 240893, 240894, 240895, 240896, 240897, 240898, 240899, 240900, 240901, 240902, 240903, 240905, 240906, 240910, 240912, 240913, 240914, 240916, 240917, 240918, 240924, 240928, 240932, 240934, 240939, 240941, 240942, 240943, 240944, 240945, 240946, 240949, 240950, 240952, 240953, 240954, 240955, 240956, 240957, 240959, 240960, 240961, 240962, 240965, 240966, 240968, 240969, 240971, 240973, 240975, 240976, 240977, 240979, 240980, 240981, 240982, 240985, 240988, 240990, 240991, 240992, 240993, 240995, 240996, 240997, 240999, 241000, 241001, 241003, 241004, 241005, 241006, 241007, 241010, 241012, 241014, 241017, 241020, 241022, 241023, 241026, 241027, 241028, 241029, 241030, 241031, 241032, 241033, 241034, 241037, 241038, 241039, 241041, 241042, 241044, 241045, 241048, 241050, 241051, 241053, 241054, 241056, 241058, 241060, 241061, 241063, 241064, 241065, 241069, 241074, 241076, 241077, 241079, 241082, 241083, 241085, 241086, 241088, 241089, 241090, 241092, 241093, 241095, 241096, 241097, 241098, 241100, 241101, 241102, 241103, 241104, 241106, 241107, 241108, 241109, 241112, 241113, 241114,
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Boston Scientific TELIGEN, Implantable Cardiac Defibrillator models E102, E110, the following models are not available in the US: models E103, E111, F102, F103, F110, F111. Sterilized using ethylene oxide. Product Usage: This family of implantable cardioverter defibrillators (ICDs) provide a variety of therapies including: -Ventricular tachyarrhythmia therapy, which is used to treat rhythms associated with sudden cardiac death such as VT and VF; -Bradycardia pacing, including adaptive rate pacing, to detect and treat bradyarrhythmias and to provide cardiac rate support after defibrillation therapy. Boston Scientific implantable cardioverter defibrillators (ICDs) are indicated to provide ventricular antitachycardia pacing (ATP) and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias recalled?

Yes. FDA published recall Z-0088-2015 on October 29, 2014. Reported hazard: Boston Scientific identified a subset of COGNISTM CRT-Ds and TELIGENTM ICDs that had experienced an increased rate of premature battery depletion due to compromised performance of a low voltage (LV) capacitor. Boston Scientific updated their August 2013 Physician communication to include additional devices that may experience premature battery depletion due to compromised performance of a low volt

What should I do if I have this product?

Voluntary: Firm initiated · Visit