Back to search
FDAFDA · Food and drugs·Recall number Z-0095-2015

The GemStar Power Supply is an accessory for the GemStar Infusion Pump. It is a two piece assembled switching power supply. The Power Supply converts alternating current (AC) line input voltage of 120 or 240 V to an output to the device of 3.3 Volts DC.

Published November 5, 2014 · Recall initiated: February 3, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The GemStar Infusion pump may not receive direct current (DC) power from the power supply. In one instance, smoke was observed and the GemStar pump was operating on battery power while connected to the 3 volt DC power supply.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Hospira Inc.
Distributor
Worldwide Distribution -- USA, including the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY; and, the countries of Austria, Belgium, Brazil, Columbia, France, Germany, Greece, Netherlands, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Turkey, and United Kingdom.
Lot
Wall Mount GemStar Power Supply List Number: 13026; Lot Numbers: 25283G1, 26286G1, 26287G1, 27292G1, 27294G1, 27295G1, 27296G1
Risk classification
Class I
Area (FDA)
device
Country
United States