Ultra Battery 1 Model: TA-UB1 NSN: 6515-01-592-5803 The Ultra Battery 1 (TA-UB1) is a rechargeable power source for the Thermal Angel Blood and IV Fluid Infusion Warmer. The TA-UB1 battery was designed to provide sufficient power for the Thermal Angel to warm and deliver one liter of blood or IV fluid infusion to a trauma patient.
Published October 29, 2014 · Recall initiated: July 25, 2014
Summary
FDA records medical device Class II recall Z-0101-2015, published October 29, 2014 and initiated July 25, 2014. Recalling firm: Estill Medical Technologies, Inc; Status: Terminated; Reported quantity: 6,886; Lot: 06-2011; 07-2011; 08-2011; 09-2011; 03-2012; 04-2012; 05-2012; 06-2012; 07-2012; 08-2012; 01-2013; 03-2013; 08-2013; 09-2013; 10-2013; 11-2013; 12-2013; 01-2014; 02-2014; 04-2014; 07-2012; 02-2014. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Estill has discovered that a fully discharged Ultra Battery 1 (TA-UB1) will reach 96.9% of its capacity during a 60 minute charging period. If you have a fully discharged Ultra Battery 1 (TA-UB1), in order to achieve a 100% of charge capacity you should charge it for a period of 90 minutes. In addition, the charging voltage range of 10-36V specified on the Ultra Battery 1 (TA-UB1) label and in t
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Estill Medical Technologies, Inc
- Distributor
- Worldwide Distribution - USA (nationwide) and Internationally to Serbia.
- Lot
- 06-2011; 07-2011; 08-2011; 09-2011; 03-2012; 04-2012; 05-2012; 06-2012; 07-2012; 08-2012; 01-2013; 03-2013; 08-2013; 09-2013; 10-2013; 11-2013; 12-2013; 01-2014; 02-2014; 04-2014; 07-2012; 02-2014
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAGYRUS ACMI, REF 3345 -Everest BiCOAG MOLly Forceps 5mm/45cm REF 3345PK - PKS MOLly Forceps 5mm/45cm REF 3640October 29, 2014
- FDASiemens ADVIA Centaur® Systems HBc Total (HBcT), Reagent. For In Vitro Diagnostic Use Catalog Number: 07566733October 29, 2014
- FDACRN Crono Syringe CRN20 with 19G x 1 "Steel Needle"October 29, 2014
- FDAADVIA Centaur®¿ Systems HBc Total (HBcT) Kit. For In Vitro Diagnostic Use. Catalog Number: 07566733 SiemensOctober 29, 2014
- FDADeRoyal(R) SafeLiner Suction Canister Kit, SEMI-RIGID, 1000cc, REF 71-6518, NON-STERILE, Rx Only, generalOctober 29, 2014
- FDAPhilips IntelliVue Monitors with software revisions J.21.03, J.21.19. Model Product MP5 M8105A; MP5SC M8105AS;October 29, 2014
Questions about this recall
Was Ultra Battery 1 Model: TA-UB1 NSN: 6515-01-592-5803 The Ultra Battery 1 (TA-UB1) is a rechargeable power source for the Thermal Angel Blood and IV Fluid Infusion Warmer. The TA-UB1 battery was designed to provide sufficient power for the Thermal Angel to warm and deliver one liter of blood or IV fluid infusion to a trauma patient. recalled?
Yes. FDA published recall Z-0101-2015 on October 29, 2014. Reported hazard: Estill has discovered that a fully discharged Ultra Battery 1 (TA-UB1) will reach 96.9% of its capacity during a 60 minute charging period. If you have a fully discharged Ultra Battery 1 (TA-UB1), in order to achieve a 100% of charge capacity you should charge it for a period of 90 minutes. In addition, the charging voltage range of 10-36V specified on the Ultra Battery 1 (TA-UB1) label and in t
What should I do if I have this product?
Voluntary: Firm initiated · Letter