AdvanDx E. coli/P. aeruginosa PNA FISH, In Vitro Diagnostic Culture Identification Kit for Escherichia coli and Pseudomonas aeruginosa Catalog Number : KT007 Usage: E. coli/P. aeruginosa PNA FISH is a multicolor, qualitative nucleic acid hybridization assay intended for identification of Escherichia coli and Pseudomonas aeruginosa on smears of positive blood cultures containing Gram-negative rods. The E.coli/P. aeruginosa PNA FISH assay is indicated for use as an aid in the diagnosis of E. coli and/or P. aeruginosa bacteremia.
Published October 31, 2012 · Recall initiated: September 14, 2012
Summary
FDA records medical device Class II recall Z-0116-2013, published October 31, 2012 and initiated September 14, 2012. Recalling firm: Advandx Inc; Status: Terminated; Reported quantity: 33 units; Lot: Lot Numbers: 01812A-US, 03412A-US and 00512A-EU Exp Date: 12-2013. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Some strains of Pseudomonas putida cross-react with the P. aeruginosa PNA probe used in E. coli/P. aeruginosa PNA FISH (KT007) due to a sequence similarity resulting in a cross-reaction that may produce a red-positive signal
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Advandx Inc
- Distributor
- Worldwide distribution: USA (nationwide) including: CA, FL, IL, MI, and NJ; and country of: Great Britain.
- Lot
- Lot Numbers: 01812A-US, 03412A-US and 00512A-EU Exp Date: 12-2013
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAdvanDx Quick FISH Fixation Kit for BC ( Blood Cultures), 50 tests. In vitro diagnostic. The QuickFISHAugust 13, 2014
- FDAAdvanDx GNR Traffic Light PNA FISH, In Vitro Diagnostic for Culture Identification Kit for EscherichiaOctober 31, 2012
- FDAGE Healthcare innova 3131-IQ Mobile X-ray system. The innova systems are indicated for use in generatingOctober 31, 2012
- FDAHiRes 90K devices with Helix Electrodes Model CI 1400-02H The HiResolution Bionic Ear System is a cochlearOctober 31, 2012
- FDAGE Healthcare innova 3100-IQ, Mobile X-ray system The innova systems are indicated for use in generatingOctober 31, 2012
- FDAOncentra External Beam 4.1. Radiation treatment planning software designed to analyze and plan radiationOctober 31, 2012
Questions about this recall
Was AdvanDx E. coli/P. aeruginosa PNA FISH, In Vitro Diagnostic Culture Identification Kit for Escherichia coli and Pseudomonas aeruginosa Catalog Number : KT007 Usage: E. coli/P. aeruginosa PNA FISH is a multicolor, qualitative nucleic acid hybridization assay intended for identification of Escherichia coli and Pseudomonas aeruginosa on smears of positive blood cultures containing Gram-negative rods. The E.coli/P. aeruginosa PNA FISH assay is indicated for use as an aid in the diagnosis of E. coli and/or P. aeruginosa bacteremia. recalled?
Yes. FDA published recall Z-0116-2013 on October 31, 2012. Reported hazard: Some strains of Pseudomonas putida cross-react with the P. aeruginosa PNA probe used in E. coli/P. aeruginosa PNA FISH (KT007) due to a sequence similarity resulting in a cross-reaction that may produce a red-positive signal
What should I do if I have this product?
Voluntary: Firm initiated · Letter