Oncentra External Beam 4.1. Radiation treatment planning software designed to analyze and plan radiation treatments in their dimension for the purpose of treating patients with cancer.

Published October 31, 2012 · Recall initiated: August 21, 2012

Summary

FDA records medical device Class II recall Z-0118-2013, published October 31, 2012 and initiated August 21, 2012. Recalling firm: Elekta, Inc.; Status: Terminated; Reported quantity: 52 units; Lot: None as the product is software. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

During the planning process the beam weights can be changed by the customers. When the beam weights are changed and renumbered and this happens during the dose calculation it is possible to save and reopen the plan without any error messages.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Elekta, Inc.
Distributor
Nationwide Distribution-including the states of AL, CA, DC, KS, LA, MO, NB, NY, NC, ND, OH, OR, PA, RI, SC, TX, and WA.
Lot
None as the product is software
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Oncentra External Beam 4.1. Radiation treatment planning software designed to analyze and plan radiation treatments in their dimension for the purpose of treating patients with cancer. recalled?

Yes. FDA published recall Z-0118-2013 on October 31, 2012. Reported hazard: During the planning process the beam weights can be changed by the customers. When the beam weights are changed and renumbered and this happens during the dose calculation it is possible to save and reopen the plan without any error messages.

What should I do if I have this product?

Voluntary: Firm initiated · Letter