FDAFDA · Food and drugs·Recall number Z-0118-2013
Oncentra External Beam 4.1. Radiation treatment planning software designed to analyze and plan radiation treatments in their dimension for the purpose of treating patients with cancer.
Published October 31, 2012 · Recall initiated: August 21, 2012
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
During the planning process the beam weights can be changed by the customers. When the beam weights are changed and renumbered and this happens during the dose calculation it is possible to save and reopen the plan without any error messages.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Elekta, Inc.
- Distributor
- Nationwide Distribution-including the states of AL, CA, DC, KS, LA, MO, NB, NY, NC, ND, OH, OR, PA, RI, SC, TX, and WA.
- Lot
- None as the product is software
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Sell products like this?