Oncentra External Beam 4.1. Radiation treatment planning software designed to analyze and plan radiation treatments in their dimension for the purpose of treating patients with cancer.
Published October 31, 2012 · Recall initiated: August 21, 2012
Summary
FDA records medical device Class II recall Z-0118-2013, published October 31, 2012 and initiated August 21, 2012. Recalling firm: Elekta, Inc.; Status: Terminated; Reported quantity: 52 units; Lot: None as the product is software. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
During the planning process the beam weights can be changed by the customers. When the beam weights are changed and renumbered and this happens during the dose calculation it is possible to save and reopen the plan without any error messages.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Elekta, Inc.
- Distributor
- Nationwide Distribution-including the states of AL, CA, DC, KS, LA, MO, NB, NY, NC, ND, OH, OR, PA, RI, SC, TX, and WA.
- Lot
- None as the product is software
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABrand Name: Geneva Product Name: Brachytherapy Applicator Product Description: The device is intended forJuly 15, 2026
- FDABrand Name: Brachytherapy Applicator Product Name: Advanced Gynecological Applicator Product Description: TheJuly 15, 2026
- FDABrand Name: Fletcher CT/MR Shielded Applicator Set Product Name: Intracavitary Remote Afterloading ApplicatorJuly 15, 2026
- FDAMOSAIQ Oncology Information SystemJune 17, 2026
- FDAHiRes 90K devices with Helix Electrodes Model CI 1400-02H The HiResolution Bionic Ear System is a cochlearOctober 31, 2012
- FDAOsseotite Tapered Certain Implant 5 X 10mm Model INT510 Biomet 3i Dental Implants are intended for surgicalOctober 31, 2012
Questions about this recall
Was Oncentra External Beam 4.1. Radiation treatment planning software designed to analyze and plan radiation treatments in their dimension for the purpose of treating patients with cancer. recalled?
Yes. FDA published recall Z-0118-2013 on October 31, 2012. Reported hazard: During the planning process the beam weights can be changed by the customers. When the beam weights are changed and renumbered and this happens during the dose calculation it is possible to save and reopen the plan without any error messages.
What should I do if I have this product?
Voluntary: Firm initiated · Letter