Back to search
FDAFDA · Food and drugs·Recall number Z-0118-2013

Oncentra External Beam 4.1. Radiation treatment planning software designed to analyze and plan radiation treatments in their dimension for the purpose of treating patients with cancer.

Published October 31, 2012 · Recall initiated: August 21, 2012

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

During the planning process the beam weights can be changed by the customers. When the beam weights are changed and renumbered and this happens during the dose calculation it is possible to save and reopen the plan without any error messages.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Elekta, Inc.
Distributor
Nationwide Distribution-including the states of AL, CA, DC, KS, LA, MO, NB, NY, NC, ND, OH, OR, PA, RI, SC, TX, and WA.
Lot
None as the product is software
Risk classification
Class II
Area (FDA)