The Centricity PACS Workstation is intended for use as a primary diagnostic and analysis tool for diagnostic images by trained healthcare professionals, including radiologists, physicians, technologists, clinicians and nurses. It is also intended for use as a clinical review workstation throughout the healthcare facility. The workstation interface provides the user with a means to display, manipulate, archive, print and export images when connected with the Centricity PACS infrastructure.
Published November 6, 2013 · Recall initiated: June 11, 2013
Summary
FDA records medical device Class II recall Z-0116-2014, published November 6, 2013 and initiated June 11, 2013. Recalling firm: GE Healthcare It; Status: Terminated; Reported quantity: 623 units; Lot: GE Centricity PACS Versions: 3.1.1.2, 3.2, 3.2.0.1, 3.2.0.2, 3.2.1, 3.2.2, 3.2.2.1, 3.2.2.2, 3.2.2.3, 3.2.6, 4.0, 4.0.1. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
GE has recently become aware of potential issues due to out of context issues associated with specific workflows of the Centricity PACS RA1000 workstation. These issues are as follows: 1) In Centricity RIS-IC, and a RIS-Driven integration, the "close" button closes the exam in RIS-IC, but does not close the exam in Centricity PACS RA1000 Workstation. 2) Interrupted workflow could break the
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Healthcare It
- Distributor
- Worldwide Distribution: US (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI and WV) and the countries of the UK, Taiwan, Switerzland, Sweden, Spain, Singapore, Saudi Arabia, Netherlands, Malaysia, Korea, Japan, Italy, Israel, India, Hong Kong, Germany, France, Denmark, China, Canada, Bermuda, Beligum and Australia.
- Lot
- GE Centricity PACS Versions: 3.1.1.2, 3.2, 3.2.0.1, 3.2.0.2, 3.2.1, 3.2.2, 3.2.2.1, 3.2.2.2, 3.2.2.3, 3.2.6, 4.0, 4.0.1
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACentricity Laboratory Core Lab System 4.1 The Centricity Laboratory System is intended to be an informationSeptember 7, 2016
- FDAThe Centricity Laboratory System is intended to be an information system designed to support the clinical andFebruary 5, 2014
- FDAMerge Mammo is a multi-modality, vendor-neutral digital mammography workstation that enables imaging centersJanuary 22, 2014
- FDAThe Centricity Laboratory System is intended to be an information system designed to support the clinical andJanuary 8, 2014
- FDABreeze C Scooter is a motorized electric scooter for elderly people. Brand name Breeze, model C. It isNovember 6, 2013
- FDABrand Names: Integra® Camino 110-4 Intracranial Pressure (ICP) Monitoring Catheters. Product / Catalogue No.s:November 6, 2013
Questions about this recall
Was The Centricity PACS Workstation is intended for use as a primary diagnostic and analysis tool for diagnostic images by trained healthcare professionals, including radiologists, physicians, technologists, clinicians and nurses. It is also intended for use as a clinical review workstation throughout the healthcare facility. The workstation interface provides the user with a means to display, manipulate, archive, print and export images when connected with the Centricity PACS infrastructure. recalled?
Yes. FDA published recall Z-0116-2014 on November 6, 2013. Reported hazard: GE has recently become aware of potential issues due to out of context issues associated with specific workflows of the Centricity PACS RA1000 workstation. These issues are as follows: 1) In Centricity RIS-IC, and a RIS-Driven integration, the "close" button closes the exam in RIS-IC, but does not close the exam in Centricity PACS RA1000 Workstation. 2) Interrupted workflow could break the
What should I do if I have this product?
Voluntary: Firm initiated · Letter