Brand Names: Integra® Camino 110-4 Intracranial Pressure (ICP) Monitoring Catheters. Product / Catalogue No.s: 110-4B, 110-4BC, 110-4G, 110-4H, 110-4HC, 110-4L, 110-4LC. The fully packaged 110-4B, 110-4H, and 110-4L catheters are also provided as a component in kits that contain other cranial access devices. These kits are Catalogue No. 110-4BC, 110- 4HC, and 110-4LC respectively. Common Name: Intracranial Pressure and Pressure-Temperature Monitoring Kits Classification Name: Device, Monitoring, Intracranial Pressure 510(k) Number: K102875 The Camino 110-4 Intracranial Pressure monitoring catheters are sterile transducer-tipped pressure monitoring catheters that are used by neurosurgeons to rapidly determine and continuously monitor intracranial pressure or pressure/temperature. The Camino 110-4B catheter is used for Intracranial Pressure Monitoring in the parenchyma or subarachnoid space. The Camino 110-4G catheter is used for Post Craniotomy Subdural Pressure Monitoring. The Camino 110-4H catheter is intended for Intracranial Pressure Monitoring and Cerebrospinal fluid drainage. The Camino 110-4L catheter is intended for use with the Licox® Brain Oxygen Monitoring System to measure intracranial pressure in the parenchyma. The fully packaged 110-4B, 110-4H, and 110-4L catheters are also provided as a component in kits that contain other cranial access devices. These kits are Catalogue No. 110-4BC, 110- 4HC, and 110-4LC respectively.
Published November 6, 2013 · Recall initiated: September 10, 2013
Summary
FDA records medical device Class II recall Z-0117-2014, published November 6, 2013 and initiated September 10, 2013. Recalling firm: Integra Limited; Status: Terminated; Reported quantity: 79,887; Lot: Product / Catalogue No.s: 110-4B, 110-4BC, 110-4G, 110-4H, 110-4HC, 110-4L, 110-4LC. Lot Numbers: Catalog Number 110-4B: 3050RX195484, 3050RX198473, 305000205426, 3050RX197641, 3050RX201421, 305000205648, 305000205652, 3050RX199149, 3050RX201407, 3050RX199850, 305000205976, 3050RY189979, 3050RX200862, 305000206178, 3050RY193648, 3050RX200486, 3050RX181026, 3050RX185115, 3050RY195484, 305000207061, 305000207062, 3050RY198473, 305000207330, 305000207334, 3050RY197641, 3050RY200862, 305000207580, 305000207892, 3050RY200486, 3050RY181026, 3050RY185115, 3050RX207330, 3050RX207334, 3050RX207580, 305000208008, 305000208510, 3050RY207892, 305000210060, 3050RX208008, 3050RX208510, 3050RX205976, 3050RX206178, 305000210563, 305000211723, 3050RX210060, 305000212616, 305000213007, 305000213191, 30500X210563, 3050RX207061, 305000214590, 3050RX205426, 30500X215502, 3050RX207062, 305000215531, 30500X216410, 3050RY199850, 305000216132, 30500X216627, 30500X217119, 3050RY217112, 305000216718, 3050RX217751, 3050RX217840, 3050RY217837, 305000217858, 30500X218369, 30500X218391, 30500X219054, 3050RY206178, 305000220708, 305000218578, 30500X220430, 305000219261, 30500X221221, 305000221226, 305000219989, 30500X222569, 305000221953, 305000221955, 30500X223153, 30500X223709, 305000223275, 305000223280, 305000222640, 30500X224142, 305000223867, 305000223866, 30500X224692, 305000224352, 305000224353, 30500X225386, 305000225045, 305000225044, 30500X225992, 305000225632, 305000225633, 30500X226540, 30500X227059, 305000226844, 305000226846, 305000226127, 30500X227564, 305000226836, 30500X227820, 305000227127, 305000227128, 30500X228194, 305000227790, 305000227789, 30500X228787, 305000228522, 305000228521, 305000228971, 305000228972, 30500X229560, 305000229328, 305000229329, 30500X229766, 30500X230087, 305000229825, 305000229826, 30500X230565, 305000230303, 305000230302, 30500X231053, 305000230780, 305000230781, 30500X231786, 305000231461, 305000231462, 30500X232331, 305000232098, 305000232099, 30500X232960, 305000232605, 305000232604, 305000233245, 305000233244, 305000233243, 305000233242, 305000233241, 305000233240, 30500X233415, 30500X234183, 305000233607, 305000233608, 30500X234806, 305000234303, 305000234304, 305000235022, 305000235017, 30500X235732, 305000235598, 30500X236328, 305000236105, 305000236104, 305000236655, 305000236653, 30500Y237087, 30500X237090, 305000237232, 30500X237761, 305000237824, 305000237825, 30500X238273, 305000238263, 305000238264, 30500X239095, 305000238818, 305000238819, 30500X239497, 30500X239748, 30500X239958, 30500X240334, 30500X240668, 30500X241368, 305000241516, 305000242236, 305000242792, 305000242795, 305000242796, 305000242794, 305000242793, 305000242791, 3050RY243869, 305000243251, 305000243882, 305000243881, 305000243888, 305000243889, 305000243883, 30500X244281, 30500X244880, 305000244513,30500X245306, 305000244481, 305000244480, 305000244934, 30500X245937, 30500X246067, 305000245408, 30500X246872, 305000245835, 30500X246951, 30500X247608, 305000246460, 30500X248284, 305000247007, 3050RX227127, 3050RX234303, 305000247717, 305000249072, 3050RX242794, 3050RX242793, 3050RX237825, 3050RX241516, 30500X249488, 305000248479, 30500X250423, 305000249743, 30500X250964, 305000250456, 30500X251060, 305000251055, 30500X251560, 30500X252133, 3050RY252136, 305000251453, 30500X252771, 305000252058, 3050RX244513, 305000252646, 305000253785, 30500X254929, 3050RY255145, 30500X255141, 305000254589, 305000255549, 305000255551, 305000255038, 30500X256614, 305000256192, 305000257281, 305000256709, 305000257280, 305000258048, 30500X259385, 305000258379, 305000258921, 305000258922, 30500X259557, 305000259431, 305000259432, 30500X260159, 305000259966, 305000259960, 305000260514, 305000260527, 30500Y261748, 3050RX253785, 305000261325, 305000261326, 305000261737, 305000261736, 3050RX256192, 30500X262441, 305000262328, 305000262327, 305000262978, 305000262977, 3050RY263508, 30500X263559, 305000263504, 305000264073, 305000264074, 3050002646. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Integra LifeSciences has identified through an internal evaluation that in order to comply with US FDA requirements, certain Camino® Intracranial Pressure Monitoring Kits require additional labelling to indicate they are UNSAFE in an MR environment. Only the following Camino® Intracranial Pressure Monitoring Kits are affected: 110-4B, 110-4BC, 110-4G, 110-4H, 110-4HC, 110-4L, 110-4LC.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Integra Limited
- Distributor
- Nationwide Distribution
- Lot
- Product / Catalogue No.s: 110-4B, 110-4BC, 110-4G, 110-4H, 110-4HC, 110-4L, 110-4LC. Lot Numbers: Catalog Number 110-4B: 3050RX195484, 3050RX198473, 305000205426, 3050RX197641, 3050RX201421, 305000205648, 305000205652, 3050RX199149, 3050RX201407, 3050RX199850, 305000205976, 3050RY189979, 3050RX200862, 305000206178, 3050RY193648, 3050RX200486, 3050RX181026, 3050RX185115, 3050RY195484, 305000207061, 305000207062, 3050RY198473, 305000207330, 305000207334, 3050RY197641, 3050RY200862, 305000207580, 305000207892, 3050RY200486, 3050RY181026, 3050RY185115, 3050RX207330, 3050RX207334, 3050RX207580, 305000208008, 305000208510, 3050RY207892, 305000210060, 3050RX208008, 3050RX208510, 3050RX205976, 3050RX206178, 305000210563, 305000211723, 3050RX210060, 305000212616, 305000213007, 305000213191, 30500X210563, 3050RX207061, 305000214590, 3050RX205426, 30500X215502, 3050RX207062, 305000215531, 30500X216410, 3050RY199850, 305000216132, 30500X216627, 30500X217119, 3050RY217112, 305000216718, 3050RX217751, 3050RX217840, 3050RY217837, 305000217858, 30500X218369, 30500X218391, 30500X219054, 3050RY206178, 305000220708, 305000218578, 30500X220430, 305000219261, 30500X221221, 305000221226, 305000219989, 30500X222569, 305000221953, 305000221955, 30500X223153, 30500X223709, 305000223275, 305000223280, 305000222640, 30500X224142, 305000223867, 305000223866, 30500X224692, 305000224352, 305000224353, 30500X225386, 305000225045, 305000225044, 30500X225992, 305000225632, 305000225633, 30500X226540, 30500X227059, 305000226844, 305000226846, 305000226127, 30500X227564, 305000226836, 30500X227820, 305000227127, 305000227128, 30500X228194, 305000227790, 305000227789, 30500X228787, 305000228522, 305000228521, 305000228971, 305000228972, 30500X229560, 305000229328, 305000229329, 30500X229766, 30500X230087, 305000229825, 305000229826, 30500X230565, 305000230303, 305000230302, 30500X231053, 305000230780, 305000230781, 30500X231786, 305000231461, 305000231462, 30500X232331, 305000232098, 305000232099, 30500X232960, 305000232605, 305000232604, 305000233245, 305000233244, 305000233243, 305000233242, 305000233241, 305000233240, 30500X233415, 30500X234183, 305000233607, 305000233608, 30500X234806, 305000234303, 305000234304, 305000235022, 305000235017, 30500X235732, 305000235598, 30500X236328, 305000236105, 305000236104, 305000236655, 305000236653, 30500Y237087, 30500X237090, 305000237232, 30500X237761, 305000237824, 305000237825, 30500X238273, 305000238263, 305000238264, 30500X239095, 305000238818, 305000238819, 30500X239497, 30500X239748, 30500X239958, 30500X240334, 30500X240668, 30500X241368, 305000241516, 305000242236, 305000242792, 305000242795, 305000242796, 305000242794, 305000242793, 305000242791, 3050RY243869, 305000243251, 305000243882, 305000243881, 305000243888, 305000243889, 305000243883, 30500X244281, 30500X244880, 305000244513,30500X245306, 305000244481, 305000244480, 305000244934, 30500X245937, 30500X246067, 305000245408, 30500X246872, 305000245835, 30500X246951, 30500X247608, 305000246460, 30500X248284, 305000247007, 3050RX227127, 3050RX234303, 305000247717, 305000249072, 3050RX242794, 3050RX242793, 3050RX237825, 3050RX241516, 30500X249488, 305000248479, 30500X250423, 305000249743, 30500X250964, 305000250456, 30500X251060, 305000251055, 30500X251560, 30500X252133, 3050RY252136, 305000251453, 30500X252771, 305000252058, 3050RX244513, 305000252646, 305000253785, 30500X254929, 3050RY255145, 30500X255141, 305000254589, 305000255549, 305000255551, 305000255038, 30500X256614, 305000256192, 305000257281, 305000256709, 305000257280, 305000258048, 30500X259385, 305000258379, 305000258921, 305000258922, 30500X259557, 305000259431, 305000259432, 30500X260159, 305000259966, 305000259960, 305000260514, 305000260527, 30500Y261748, 3050RX253785, 305000261325, 305000261326, 305000261737, 305000261736, 3050RX256192, 30500X262441, 305000262328, 305000262327, 305000262978, 305000262977, 3050RY263508, 30500X263559, 305000263504, 305000264073, 305000264074, 3050002646
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAThe Centricity PACS Workstation is intended for use as a primary diagnostic and analysis tool for diagnosticNovember 6, 2013
- FDACentricity PACS IW by GE Healthcare Dynamic Imaging Solutions is a device that receives medical imagesNovember 6, 2013
- FDAIntegra Leyla Ball Joint Clamp, also as included in Leyla Table Mounting Hardware. The Leyla RetractionSeptember 4, 2013
- FDAThe MAYFIELD® Horseshoe Gel Pads are used in conjunction with the MAYFIELD® Base Units and MAYFIELD® SwivelOctober 3, 2012
- FDAThe MAYFIELD® Horseshoe Gel Pads are used in conjunction with the MAYFIELD® Base Units and MAYFIELD® SwivelOctober 3, 2012
- FDAIntegra(TM) Jarit(R) Take-Apart Endoscopic Instruments. Integra(TM) Jarit(R) Take-Apart Endoscopic InstrumentsJuly 25, 2012
Questions about this recall
Was Brand Names: Integra® Camino 110-4 Intracranial Pressure (ICP) Monitoring Catheters. Product / Catalogue No.s: 110-4B, 110-4BC, 110-4G, 110-4H, 110-4HC, 110-4L, 110-4LC. The fully packaged 110-4B, 110-4H, and 110-4L catheters are also provided as a component in kits that contain other cranial access devices. These kits are Catalogue No. 110-4BC, 110- 4HC, and 110-4LC respectively. Common Name: Intracranial Pressure and Pressure-Temperature Monitoring Kits Classification Name: Device, Monitoring, Intracranial Pressure 510(k) Number: K102875 The Camino 110-4 Intracranial Pressure monitoring catheters are sterile transducer-tipped pressure monitoring catheters that are used by neurosurgeons to rapidly determine and continuously monitor intracranial pressure or pressure/temperature. The Camino 110-4B catheter is used for Intracranial Pressure Monitoring in the parenchyma or subarachnoid space. The Camino 110-4G catheter is used for Post Craniotomy Subdural Pressure Monitoring. The Camino 110-4H catheter is intended for Intracranial Pressure Monitoring and Cerebrospinal fluid drainage. The Camino 110-4L catheter is intended for use with the Licox® Brain Oxygen Monitoring System to measure intracranial pressure in the parenchyma. The fully packaged 110-4B, 110-4H, and 110-4L catheters are also provided as a component in kits that contain other cranial access devices. These kits are Catalogue No. 110-4BC, 110- 4HC, and 110-4LC respectively. recalled?
Yes. FDA published recall Z-0117-2014 on November 6, 2013. Reported hazard: Integra LifeSciences has identified through an internal evaluation that in order to comply with US FDA requirements, certain Camino® Intracranial Pressure Monitoring Kits require additional labelling to indicate they are UNSAFE in an MR environment. Only the following Camino® Intracranial Pressure Monitoring Kits are affected: 110-4B, 110-4BC, 110-4G, 110-4H, 110-4HC, 110-4L, 110-4LC.
What should I do if I have this product?
Voluntary: Firm initiated · Letter