FDAFDA · Food and drugs·Recall number Z-0133-2027

V*Home, Model Number: PRT-01444-000, Package PN:PRT-01445-000. used for ventilatory support.

Published October 28, 2026 · Recall initiated: September 17, 2026

Summary

FDA records medical device Class I recall Z-0133-2027, published October 28, 2026 and initiated September 17, 2026. Recalling firm: Ventec Life Systems, Inc.; Status: Ongoing; Reported quantity: 25,112 units; Lot: All Serial Numbers: UDI: 00855573007822. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Ventec Life Systems, Inc.
Distributor
Distribution US nationwide including Puerto Rico.
Lot
All Serial Numbers: UDI: 00855573007822
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was V*Home, Model Number: PRT-01444-000, Package PN:PRT-01445-000. used for ventilatory support. recalled?

Yes. FDA published recall Z-0133-2027 on October 28, 2026. Reported hazard: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.

What should I do if I have this product?

Voluntary: Firm initiated · Letter