Philips Lumify Diagnostic Ultrasound System
Published October 28, 2026 · Recall initiated: September 24, 2026
Summary
FDA records medical device Class I recall Z-0152-2027, published October 28, 2026 and initiated September 24, 2026. Recalling firm: Philips Ultrasound, LLC; Status: Ongoing; Reported quantity: 13,206; Lot: UDI-DI/Software Version: Android: 00884838122116/1.X, 3.X, 4.X, 5.X. iOS: 00884838124141/2.X and 5.X. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Ultrasound system may: 1) Not recognize previous transducer registration, prevent scanning, require re-registration with Wi-Fi/cellular access, which may be unavailable outside of the hospital (ambulances/aircraft/field); 2) At connection between transducer USB cable to tablet/phone, there may be an unstable connection/may disconnect disrupting function. May cause diagnosis/treatment delay.
What to do
Voluntary: Firm initiated
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Ultrasound, LLC
- Distributor
- Worldwide - US Nationwide distribution including in the states of CA, MA, IL, NY, OR, GA, TX, WA, FL, PA, WI, LA, NJ, KY, VA, MS, AZ, SC, CT, MN, UT, OH, AL, MI, NH, MO, TN, IN, DC, NE, NM, MT, NV, NC, MD, IA, SD, VT, KS, AK, CO, ID, OK, WV, AR, RI, ME, WY, HI, DE, ND and the countries of CA, NO, HR, DE, KR, GB, ES, FI, PL, AU, TH, SG, NL, RO, CH, FR, MM, MX, BE, DK, MY, SE, LB, SI, BR, CL, GR, PH, CN, ID, LT, HK, MQ, JP, IT, IE, IL, AT, TW, AE, SA, HU, RU, VN, NZ, TR, SK, ZA, KE, AR, NP, PK, UY, DO, MN, LK, GT, BD, IS, IN, NI, LY, CY, CZ, SV, PA, BG, PE, EC, CO, VE, UZ, YE, KW, DZ, LU, EG, EE, UA, LV, KZ, TZ, MD, JO, GY, QA, MU, AM, OM, PT, TN, NG, BO, PF, ZM, GF, YT, ME, GD, RS, BW, GP, RE, MT, CR, MA, CI, NC, BA, SR, CG, TL, GH, AL, MV, XK, BN, LS, MK, CM, TT, ZW, BH, PS, HN, MC, WS, JE, PM, AZ.
- Lot
- UDI-DI/Software Version: Android: 00884838122116/1.X, 3.X, 4.X, 5.X. iOS: 00884838124141/2.X and 5.X
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAV*Home, Model Number: PRT-01444-000, Package PN:PRT-01445-000. used for ventilatory support.October 28, 2026
- FDAAmplatzer Amulet Left Atrial Appendage Occluder Model Numbers 1. 9-ACP2-007-016 2. 9-ACP2-007-018 3.October 28, 2026
- FDAV+ Pro Emergency, Model Number: PRT-01201-000, Package PN:PRT-01173-000,PRT-01222-000. used for ventilatoryOctober 28, 2026
- FDAThink Surgical Square Driver REF: 109126October 28, 2026
- FDAVC(Japan) Model Number: PRT-01098-100, Package PN:PRT-01125-100. used for ventilatory support.October 28, 2026
- FDAPhilips EPIQ Elite Diagnostic Ultrasound System Model # 795234September 30, 2026
Questions about this recall
Was Philips Lumify Diagnostic Ultrasound System recalled?
Yes. FDA published recall Z-0152-2027 on October 28, 2026. Reported hazard: Ultrasound system may: 1) Not recognize previous transducer registration, prevent scanning, require re-registration with Wi-Fi/cellular access, which may be unavailable outside of the hospital (ambulances/aircraft/field); 2) At connection between transducer USB cable to tablet/phone, there may be an unstable connection/may disconnect disrupting function. May cause diagnosis/treatment delay.
What should I do if I have this product?
Voluntary: Firm initiated