6.5mm Cancellous Bone Screw Intended to be used for the fixation of the acetabular shell or fracture repair.
Published October 31, 2012 · Recall initiated: September 28, 2012
Summary
FDA records medical device Class II recall Z-0139-2013, published October 31, 2012 and initiated September 28, 2012. Recalling firm: Encore Medical, Lp; Status: Terminated; Reported quantity: 1 unit; Lot: Model Number 10-55-025, Lot Number 006A1057. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A product complaint was received where a size 20 screw was labeled as size 30. After receipt, measurement of the screw determined it to be a size 25. Further investigation also determined that there were other screws labeled as size 25, but a size 30 screw was inside of the package. It was thought that all of these screws were found in-house and were contained, but it was determined later that on
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Encore Medical, Lp
- Distributor
- Nationwide Distribution only to CA.
- Lot
- Model Number 10-55-025, Lot Number 006A1057.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAltivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 54mm, REF 804-07-540 Product Usage: TheApril 4, 2018
- FDAAltivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 50mm, REF 804-07-500 Product Usage: TheApril 4, 2018
- FDACobalt HV with Gentamicin, Cobalt Bone Cement 40GM, REF 402283, QTY 1, STERILE, Rx Only, Mfgr: BIOMETNovember 15, 2017
- FDACobalt MV Bone Cement 40GM, REF 402438, QTY 1, STERILE, Rx Only, Mfgr: BIOMET ORTHOPEDICS Cobalt MV and HVNovember 15, 2017
- FDAAquaLiner Hydrophilic Guidewire 0.035" X 180cm, Angle / Regular. AquaLiner Hydrophilic Guidewire is designedOctober 31, 2012
- FDASpiraLok 5.0mm with Ethibond and needles Product Number: 222960 SPIRALOK Anchors are primarily used toOctober 31, 2012
Questions about this recall
Was 6.5mm Cancellous Bone Screw Intended to be used for the fixation of the acetabular shell or fracture repair. recalled?
Yes. FDA published recall Z-0139-2013 on October 31, 2012. Reported hazard: A product complaint was received where a size 20 screw was labeled as size 30. After receipt, measurement of the screw determined it to be a size 25. Further investigation also determined that there were other screws labeled as size 25, but a size 30 screw was inside of the package. It was thought that all of these screws were found in-house and were contained, but it was determined later that on
What should I do if I have this product?
Voluntary: Firm initiated · Letter