Harmony Equipment Management System, Model #603X/80XX, STERIS Corporation, Mentor, OH Harmony Equipment Management Systems (EMS) are used in hospital facilities to provide ready access to medical gasses, electrical power and audio/visual data services, in addition to managing surgical support equipment such as endoscopic equipment, cautery devices and insufflators.
Published November 13, 2013 · Recall initiated: August 15, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Dr¿ger Medical GmbH has notified STERIS of the potential for a break in the drive screw contained in the motorized ceiling supply unit, a component of the Harmony EMS system. The ceiling supply unit controls the motorized adjustment of the height and rotation of the supply column. Should a break occur in the drive screw, there is potential for the adjustable arm to drop down, causing the EMS sys
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Steris Corporation
- Distributor
- Nationwide distribution: USA including the states of Alabama, Arkansas, Florida, Georgia, Illinois, Indiana, Kansas, Kentucky, Louisiana, Maryland, Massachusetts, Michigan, Minnesota, Missouri, Nebraska, New Jersey, New York, Ohio, Oklahoma, Oregon, Pennsylvania, Texas, Virginia, Washington and Wisconsin.
- Lot
- Serial numbers 0422503114 - 0426604096
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States