GE Healthcare Giraffe Incubator Carestation CS1, Infant Incubator. Model Numbers: 1) 2082844-002-01040417; 2) 2082844-002-01044956; 3) 2082844-002-01058424; 4) 2082844-002-01069017; 5) 2082844-002-01089666; 6) 2082844-002-01090474; 7) 2082844-002-01091269; 8) 2082844-002-01093489; 9) 2082844-002-01113262; 10) 2082844-002-01113711; 11) 2082844-002-01123085; 12) 2082844-002-01123918; 13) 2082844-002-01128966; 14) 2082844-002-01133042; 15) 2082844-002-01146447; 16) 2082844-002-100205; 17) 2082844-002-100749; 18) 2082844-002-100752; 19) 2082844-002-100995; 20) 2082844-002-101425; 21) 2082844-002-523548; 22) 2082844-002-529928; 23) 2082844-002-594800; 24) 2082844-002-643544; 25) 2082844-002-925841; 26) 2082844-002-965540; 27) 2082844-002-982916; 28) 2082844-002-983420; 29) 2082844-002-993905; 30) 2082844-002-995018; 31) 2082844-002-995799; 32) 2082844-002-998943; Radiant infant warmer

Published November 6, 2024 · Recall initiated: June 6, 2023

Summary

FDA records medical device Class I recall Z-0141-2025, published November 6, 2024 and initiated June 6, 2023. Recalling firm: DATEX--OHMEDA, INC.; Status: Completed; Reported quantity: 224 units; Lot: 1) 2082844-002-01040417, UDI: (01)00840682116855(21)TAAC70003(11)230106, (01)00840682116855(21)TAAC70004(11)230109, (01)00840682116855(21)TAAC70011(11)230110, (01)00840682116855(21)TAAC70012(11)230110, (01)00840682116855(21)TAAC70013(11)230111, (01)00840682116855(21)TAAC70015(11)230112, (01)00840682116855(21)TAAC70016(11)230112, (01)00840682116855(21)TAAC70020(11)230112, (01)00840682116855(21)TAAC70022(11)230112, (01)00840682116855(21)TAAC70024(11)230112, (01)00840682116855(21)TAAC70147(11)230317, (01)00840682116855(21)TAAC70148(11)230317, (01)00840682116855(21)TAAC70149(11)230317, (01)00840682116855(21)TAAC70150(11)230317, (01)00840682116855(21)TAAC70152(11)230320, (01)00840682116855(21)TAAC70153(11)230320, (01)00840682116855(21)TAAC70155(11)230322, (01)00840682116855(21)TAAC70158(11)230322, (01)00840682116855(21)TAAC70159(11)230323, (01)00840682116855(21)TAAC70160(11)230323, (01)00840682116855(21)TAAC70161(11)230323, (01)00840682116855(21)TAAC70162(11)230323, (01)00840682116855(21)TAAC70163(11)230324, (01)00840682116855(21)TAAC70164(11)230324, (01)00840682116855(21)TAAC70165(11)230324, (01)00840682116855(21)TAAC70166(11)230324, (01)00840682116855(21)TAAC70167(11)230327, (01)00840682116855(21)TAAC70169(11)230327, (01)00840682116855(21)TAAC70170(11)230327, (01)00840682116855(21)TAAC70171(11)230328, (01)00840682116855(21)TAAC70177(11)230330, (01)00840682116855(21)TAAC70178(11)230330, (01)00840682116855(21)TAAC70179(11)230331, (01)00840682116855(21)TAAC70181(11)230331, (01)00840682116855(21)TAAC70182(11)230331, (01)00840682116855(21)TAAC70184(11)230331, (01)00840682116855(21)TAAC70192(11)230405, (01)00840682116855(21)TAAC70193(11)230405, (01)00840682116855(21)TAAC70196(11)230405, (01)00840682116855(21)TAAC70197(11)230405, (01)00840682116855(21)TAAC70200(11)230407, (01)00840682116855(21)TAAC70201(11)230407, (01)00840682116855(21)TAAC70202(11)230407, (01)00840682116855(21)TAAC70203(11)230407, (01)00840682116855(21)TAAC70204(11)230407, (01)00840682116855(21)TAAC70205(11)230407, (01)00840682116855(21)TAAC70208(11)230411, (01)00840682116855(21)TAAC70209(11)230411, (01)00840682116855(21)TAAC70210(11)230411, (01)00840682116855(21)TAAC70211(11)230414, (01)00840682116855(21)TAAC70212(11)230414, (01)00840682116855(21)TAAC70213(11)230414, (01)00840682116855(21)TAAC70214(11)230414, (01)00840682116855(21)TAAC70215(11)230417, (01)00840682116855(21)TAAC70216(11)230417, (01)00840682116855(21)TAAC70217(11)230417, (01)00840682116855(21)TAAC70218(11)230417, (01)00840682116855(21)TAAC70223(11)230418, (01)00840682116855(21)TAAC70224(11)230418, (01)00840682116855(21)TAAC70225(11)230418, (01)00840682116855(21)TAAC70226(11)230418, (01)00840682116855(21)TAAC70227(11)230418, (01)00840682116855(21)TAAC70228(11)230418, (01)00840682116855(21)TAAC70229(11)230419, (01)00840682116855(21)TAAC70242(11)230421, (01)00840682116855(21)TAAC70243(11)230421, (01)00840682116855(21)TAAC70244(11)230421, (01)00840682116855(21)TAAC70245(11)230421, (01)00840682116855(21)TAAC70246(11)230421, (01)00840682116855(21)TAAC70247(11)230421; 2) 2082844-002-01044956, UDI: (01)008406821168552(21)TAA23140007SA(11)230404, (01)008406821168552(21)TAA23140008SA(11)230404; 3) 2082844-002-01058424, UDI: (01)00840682116855(21)TAAC70142(11)230213; 4) 2082844-002-01069017, UDI: (01)008406821168552(21)TAA23110001SA(11)230316, (01)008406821168552(21)TAA23110002SA(11)230316, (01)008406821168552(21)TAA23110003SA(11)230316; 5) 2082844-002-01089666, UDI: (01)00840682116855(21)TAAC70124(11)230201; 6) 2082844-002-01090474, UDI: (01)00840682116855(21)TAAC70126(11)230201, (01)00840682116855(21)TAAC70128(11)230206, (01)00840682116862(21)TAAC70132(11)230206, (01)00840682116855(21)TAAC70133(11)230206, (01)00840682116855(21)TAAC70125(11)230201, (01)00840682116855(21)TAAC70127(11)230206, (01)00840682116855(21)TAAC70129(11)230206, (01)00840682116855(21)TAAC70131(11)230206, (01)00840682116855(21)TAAC70136(11)230206, (01)00840682116855(21)TAAC70137(11)230206; 7) 2082844-002-01091269, UDI: (01)0. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Completed — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Prior to current established manufacturing practices, higher levels of airborne formaldehyde exposure from newly manufactured incubators could be seen during clinical use that would rapidly decrease over a week. Affected customers were notified in June 2023 to run incubators for a week to potentially off gas any residual airborne formaldehyde. The general scientific consensus is that sensory irritation of the eyes is the most sensitive endpoint with respect to exposure to airborne formaldehyde and is the adverse effect clinically experienced at the lowest concentration.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
DATEX--OHMEDA, INC.
Distributor
worldwide
Lot
1) 2082844-002-01040417, UDI: (01)00840682116855(21)TAAC70003(11)230106, (01)00840682116855(21)TAAC70004(11)230109, (01)00840682116855(21)TAAC70011(11)230110, (01)00840682116855(21)TAAC70012(11)230110, (01)00840682116855(21)TAAC70013(11)230111, (01)00840682116855(21)TAAC70015(11)230112, (01)00840682116855(21)TAAC70016(11)230112, (01)00840682116855(21)TAAC70020(11)230112, (01)00840682116855(21)TAAC70022(11)230112, (01)00840682116855(21)TAAC70024(11)230112, (01)00840682116855(21)TAAC70147(11)230317, (01)00840682116855(21)TAAC70148(11)230317, (01)00840682116855(21)TAAC70149(11)230317, (01)00840682116855(21)TAAC70150(11)230317, (01)00840682116855(21)TAAC70152(11)230320, (01)00840682116855(21)TAAC70153(11)230320, (01)00840682116855(21)TAAC70155(11)230322, (01)00840682116855(21)TAAC70158(11)230322, (01)00840682116855(21)TAAC70159(11)230323, (01)00840682116855(21)TAAC70160(11)230323, (01)00840682116855(21)TAAC70161(11)230323, (01)00840682116855(21)TAAC70162(11)230323, (01)00840682116855(21)TAAC70163(11)230324, (01)00840682116855(21)TAAC70164(11)230324, (01)00840682116855(21)TAAC70165(11)230324, (01)00840682116855(21)TAAC70166(11)230324, (01)00840682116855(21)TAAC70167(11)230327, (01)00840682116855(21)TAAC70169(11)230327, (01)00840682116855(21)TAAC70170(11)230327, (01)00840682116855(21)TAAC70171(11)230328, (01)00840682116855(21)TAAC70177(11)230330, (01)00840682116855(21)TAAC70178(11)230330, (01)00840682116855(21)TAAC70179(11)230331, (01)00840682116855(21)TAAC70181(11)230331, (01)00840682116855(21)TAAC70182(11)230331, (01)00840682116855(21)TAAC70184(11)230331, (01)00840682116855(21)TAAC70192(11)230405, (01)00840682116855(21)TAAC70193(11)230405, (01)00840682116855(21)TAAC70196(11)230405, (01)00840682116855(21)TAAC70197(11)230405, (01)00840682116855(21)TAAC70200(11)230407, (01)00840682116855(21)TAAC70201(11)230407, (01)00840682116855(21)TAAC70202(11)230407, (01)00840682116855(21)TAAC70203(11)230407, (01)00840682116855(21)TAAC70204(11)230407, (01)00840682116855(21)TAAC70205(11)230407, (01)00840682116855(21)TAAC70208(11)230411, (01)00840682116855(21)TAAC70209(11)230411, (01)00840682116855(21)TAAC70210(11)230411, (01)00840682116855(21)TAAC70211(11)230414, (01)00840682116855(21)TAAC70212(11)230414, (01)00840682116855(21)TAAC70213(11)230414, (01)00840682116855(21)TAAC70214(11)230414, (01)00840682116855(21)TAAC70215(11)230417, (01)00840682116855(21)TAAC70216(11)230417, (01)00840682116855(21)TAAC70217(11)230417, (01)00840682116855(21)TAAC70218(11)230417, (01)00840682116855(21)TAAC70223(11)230418, (01)00840682116855(21)TAAC70224(11)230418, (01)00840682116855(21)TAAC70225(11)230418, (01)00840682116855(21)TAAC70226(11)230418, (01)00840682116855(21)TAAC70227(11)230418, (01)00840682116855(21)TAAC70228(11)230418, (01)00840682116855(21)TAAC70229(11)230419, (01)00840682116855(21)TAAC70242(11)230421, (01)00840682116855(21)TAAC70243(11)230421, (01)00840682116855(21)TAAC70244(11)230421, (01)00840682116855(21)TAAC70245(11)230421, (01)00840682116855(21)TAAC70246(11)230421, (01)00840682116855(21)TAAC70247(11)230421; 2) 2082844-002-01044956, UDI: (01)008406821168552(21)TAA23140007SA(11)230404, (01)008406821168552(21)TAA23140008SA(11)230404; 3) 2082844-002-01058424, UDI: (01)00840682116855(21)TAAC70142(11)230213; 4) 2082844-002-01069017, UDI: (01)008406821168552(21)TAA23110001SA(11)230316, (01)008406821168552(21)TAA23110002SA(11)230316, (01)008406821168552(21)TAA23110003SA(11)230316; 5) 2082844-002-01089666, UDI: (01)00840682116855(21)TAAC70124(11)230201; 6) 2082844-002-01090474, UDI: (01)00840682116855(21)TAAC70126(11)230201, (01)00840682116855(21)TAAC70128(11)230206, (01)00840682116862(21)TAAC70132(11)230206, (01)00840682116855(21)TAAC70133(11)230206, (01)00840682116855(21)TAAC70125(11)230201, (01)00840682116855(21)TAAC70127(11)230206, (01)00840682116855(21)TAAC70129(11)230206, (01)00840682116855(21)TAAC70131(11)230206, (01)00840682116855(21)TAAC70136(11)230206, (01)00840682116855(21)TAAC70137(11)230206; 7) 2082844-002-01091269, UDI: (01)0
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was GE Healthcare Giraffe Incubator Carestation CS1, Infant Incubator. Model Numbers: 1) 2082844-002-01040417; 2) 2082844-002-01044956; 3) 2082844-002-01058424; 4) 2082844-002-01069017; 5) 2082844-002-01089666; 6) 2082844-002-01090474; 7) 2082844-002-01091269; 8) 2082844-002-01093489; 9) 2082844-002-01113262; 10) 2082844-002-01113711; 11) 2082844-002-01123085; 12) 2082844-002-01123918; 13) 2082844-002-01128966; 14) 2082844-002-01133042; 15) 2082844-002-01146447; 16) 2082844-002-100205; 17) 2082844-002-100749; 18) 2082844-002-100752; 19) 2082844-002-100995; 20) 2082844-002-101425; 21) 2082844-002-523548; 22) 2082844-002-529928; 23) 2082844-002-594800; 24) 2082844-002-643544; 25) 2082844-002-925841; 26) 2082844-002-965540; 27) 2082844-002-982916; 28) 2082844-002-983420; 29) 2082844-002-993905; 30) 2082844-002-995018; 31) 2082844-002-995799; 32) 2082844-002-998943; Radiant infant warmer recalled?

Yes. FDA published recall Z-0141-2025 on November 6, 2024. Reported hazard: Prior to current established manufacturing practices, higher levels of airborne formaldehyde exposure from newly manufactured incubators could be seen during clinical use that would rapidly decrease over a week. Affected customers were notified in June 2023 to run incubators for a week to potentially off gas any residual airborne formaldehyde. The general scientific consensus is that sensory irritation of the eyes is the most sensitive endpoint with respect to exposure to airborne formaldehyde and is the adverse effect clinically experienced at the lowest concentration.

What should I do if I have this product?

Voluntary: Firm initiated · Letter