Phasitron5 Breathing Circuit (P5-10, P5-HC, P5-UC-10). component of Intrapulmonary Percussive Ventilation, used to provide airway clearance therapy
Published November 6, 2024 · Recall initiated: September 16, 2024
Summary
FDA records medical device Class I recall Z-0208-2025, published November 6, 2024 and initiated September 16, 2024. Recalling firm: Percussionaire Corporation; Status: Ongoing; Reported quantity: 43,830 devices; Lot: Model Numbers: P5-10 P5-HC P5-UC-10 UDI-DI codes: Lot Numbers: 240321 240329 240412 240430 240509 240516 240614 240625 240627. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
As a result of product complaints, breathing circuits are experiencing discolored fluid which are formed when therapy was being given with specific medication through the nebulizer cup. Discoloration is caused by a reaction between medicated mist and an inadvertent/unapproved addition of a nickel coating on the spring component.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Percussionaire Corporation
- Distributor
- U.S.: PA, AR, MI, CA, TX, OH, NY, FL, IL, NE, OR, GA, WI, MO, OK, AL, WV, MT, CO, LA, VA, CT, SC, NH, NV, NC, NJ, MI, UT, ID, ME, MN, SD, ID, ND, KS, KY, TN, MD, VT, MA, RI, O.U.S.: Canada, Switzerland (all products shipped OUS remained in Sentec AG's (distributor, parent company of Percussionaire) control, were quarantined, and did not leave the warehouse).
- Lot
- Model Numbers: P5-10 P5-HC P5-UC-10 UDI-DI codes: Lot Numbers: 240321 240329 240412 240430 240509 240516 240614 240625 240627
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAIntrapulmonary Percussive Ventilation, IPV-2C, Product Number: S00002-C; Phasitron 5, Product Number: P5 usedOctober 7, 2026
- FDAImpulsator, Product Number: S00010, Phasitron 5, Product Number: P5 used with Instruction ManualOctober 7, 2026
- FDAIntrapulmonary Percussive Ventilation, IPV-1C, Product Number: S00002-C; Phasitron 5, Product Number: P5 usedOctober 7, 2026
- FDAIn-Line Valve, Product Number: P5-TEEOctober 7, 2026
- FDAGE Healthcare Giraffe Incubator Carestation CS1, Infant Incubator. Model Numbers: 1) 2082844-002-01040417; 2)November 6, 2024
- FDACADD-Solis Li-ion Rechargeable Battery Packs, REF 21-2160-XX, which are also sold as a component of theNovember 6, 2024
Questions about this recall
Was Phasitron5 Breathing Circuit (P5-10, P5-HC, P5-UC-10). component of Intrapulmonary Percussive Ventilation, used to provide airway clearance therapy recalled?
Yes. FDA published recall Z-0208-2025 on November 6, 2024. Reported hazard: As a result of product complaints, breathing circuits are experiencing discolored fluid which are formed when therapy was being given with specific medication through the nebulizer cup. Discoloration is caused by a reaction between medicated mist and an inadvertent/unapproved addition of a nickel coating on the spring component.
What should I do if I have this product?
Voluntary: Firm initiated · Letter