Back to search
FDAFDA · Food and drugs·Recall number Z-0159-2027

MAS Omni IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, REF: OPRO-101; MAS Omni IMMUNETM PRO Liquid Assayed Integrated Immunoassay Control, Sample Pack, REF: OPRO-SP

Published October 21, 2026 · Recall initiated: August 17, 2026

Summary

FDA records medical device Class II recall Z-0159-2027, published October 21, 2026 and initiated August 17, 2026. Recalling firm: Microgenics Corporation; Status: Ongoing; Reported quantity: 1366; Lot: REF/UDI-DI/Lot(Expiration): OPRO-101/00884883005327/OPRO29061A(2029-06-30); OPRO-SP/00884883005358/OPRO2906SA(2029-06-30). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Vial-to-vial variability in MAS® Omni"IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, Lot OPRO29061A results in low Parathyroid Hormone (PTH) recovery, which may lead to a delay in reporting patient PTH results.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Microgenics Corporation
Distributor
Worldwide - US Nationwide distribution in the states of UT, MI, OK, NY, TX, CA, LA, SC, OH, MO, OR and the countries of KOR, BGD, THA, ZAF, DEU, VNM, BRA, NZL.
Lot
REF/UDI-DI/Lot(Expiration): OPRO-101/00884883005327/OPRO29061A(2029-06-30); OPRO-SP/00884883005358/OPRO2906SA(2029-06-30)
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was MAS Omni IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, REF: OPRO-101; MAS Omni IMMUNETM PRO Liquid Assayed Integrated Immunoassay Control, Sample Pack, REF: OPRO-SP recalled?

Yes. FDA published recall Z-0159-2027 on October 21, 2026. Reported hazard: Vial-to-vial variability in MAS® Omni"IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, Lot OPRO29061A results in low Parathyroid Hormone (PTH) recovery, which may lead to a delay in reporting patient PTH results.

What should I do if I have this product?

Voluntary: Firm initiated · Letter