MAS Omni Immune PRO Controls
Published July 2, 2014 · Recall initiated: December 2, 2013
Summary
FDA records medical device Class III recall Z-1930-2014, published July 2, 2014 and initiated December 2, 2013. Recalling firm: Microgenics Corporation; Status: Terminated; Reported quantity: 1393; Lot: Part Number Lot number OPR0-101 OPR015101 OPR0-202 OPR015102 OPR0-303 OPR015103 OPRO-SP OPR01510S All lots have an expiration date of 10/31/2015. Class III means the product is not likely to cause adverse health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The constituent, Inhibin A, was removed from MAS Omni IMMUNE and MAS Omni IMMUNE PRO controls; however the package insert was not updated to reflect the change.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Microgenics Corporation
- Distributor
- Worldwide Distribution-USA (nationwide) and the countries of Australia, China, Germany, Hong Kong and Malaysia.
- Lot
- Part Number Lot number OPR0-101 OPR015101 OPR0-202 OPR015102 OPR0-303 OPR015103 OPRO-SP OPR01510S All lots have an expiration date of 10/31/2015.
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMAS Omni IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, REF: OPRO-101; MAS Omni IMMUNETMOctober 21, 2026
- FDAMAS CardioImmune XL Cardiac Marker Control; Model: CAI-XL1; intended for use in the clinical laboratory as anDecember 10, 2014
- FDABystolic (nebivolol) tablets, 20 mg/tablet. 7 Tablet Professional Sample Bottle, Rx Only. ForestJuly 9, 2014
- FDAMAS Omni Immune ControlsJuly 2, 2014
- FDAAlere Triage BNP Calibrators for the Beckman Coulter Access Family of Immunoassay Systems, Model No. 98202 TheJuly 2, 2014
- FDAThermo Scientific MAS Bilirubin for Siemens and Dimension Vista Systems, , Product is manufactured andJuly 3, 2013
Questions about this recall
Was MAS Omni Immune PRO Controls recalled?
Yes. FDA published recall Z-1930-2014 on July 2, 2014. Reported hazard: The constituent, Inhibin A, was removed from MAS Omni IMMUNE and MAS Omni IMMUNE PRO controls; however the package insert was not updated to reflect the change.
What should I do if I have this product?
Voluntary: Firm initiated · Letter