Williams Cystoscopic Injection Needles, used for cystoscopic injection into the urethra, bladder neck, and bladder wall, Reference Part Number 090001-S26, Order Number (GPN) G16112

Published October 21, 2026 · Recall initiated: September 16, 2026

Summary

FDA records medical device Class II recall Z-0163-2027, published October 21, 2026 and initiated September 16, 2026. Recalling firm: Cook Incorporated; Status: Ongoing; Reported quantity: 97 units; Lot: UDI-DI: 00827002161122; Lot Numbers: 16613606, 16951548, 16951548X. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The female Luer lock adapter component in the affected device lots may not have been correctly seated due to misalignment and inadequate torquing during manufacturing. Potential events that may occur as a result of this issue include increased procedural time and inadequate treatment.

What to do

Voluntary: Firm initiated

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Cook Incorporated
Distributor
US States: FL, IL, NM, TX, WA. Canada, France, Germany, Italy, Spain, Switzerland.
Lot
UDI-DI: 00827002161122; Lot Numbers: 16613606, 16951548, 16951548X
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Williams Cystoscopic Injection Needles, used for cystoscopic injection into the urethra, bladder neck, and bladder wall, Reference Part Number 090001-S26, Order Number (GPN) G16112 recalled?

Yes. FDA published recall Z-0163-2027 on October 21, 2026. Reported hazard: The female Luer lock adapter component in the affected device lots may not have been correctly seated due to misalignment and inadequate torquing during manufacturing. Potential events that may occur as a result of this issue include increased procedural time and inadequate treatment.

What should I do if I have this product?

Voluntary: Firm initiated