TS Universal Baseplate Finishing Tray, Triathlon. Catalog Number: 6543-6-506.

Published October 21, 2026 · Recall initiated: September 16, 2026

Summary

FDA records medical device Class II recall Z-0164-2027, published October 21, 2026 and initiated September 16, 2026. Recalling firm: Howmedica Osteonics Corp.; Status: Ongoing; Reported quantity: 605 trays; Lot: Catalog Number: 6543-6-506. GTIN: 07613327548082. All lot numbers used prior to the correction date of May 8, 2026 are impacted. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Howmedica Osteonics Corp.
Distributor
These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.
Lot
Catalog Number: 6543-6-506. GTIN: 07613327548082. All lot numbers used prior to the correction date of May 8, 2026 are impacted.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was TS Universal Baseplate Finishing Tray, Triathlon. Catalog Number: 6543-6-506. recalled?

Yes. FDA published recall Z-0164-2027 on October 21, 2026. Reported hazard: Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.

What should I do if I have this product?

Voluntary: Firm initiated · Letter