Back to search
FDAFDA · Food and drugs·Recall number Z-0167-2013

SCP Pump Control Panel, Product Code: 60-02-50, Serial Numbers: 60S09131, 60S09133-60S09135, 60S09140-60S09142, 60S08954. The SCP is a cardiopulmonary bypass speed control device indicated for use exclusively with the COBE Revolution pump head for speed-controlled pumping through cardiopulmonary bypass circuit for typical durations of six hours or less.

Published November 7, 2012 · Recall initiated: September 20, 2012

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Stockert Centrifugal Pump (SCP) Pump Control Panel . An out of specification circuit board component within the Panel may lead to a malfunction consisting of a blank or flashing display. The flow and pressure display information may also be incorrect after the malfunction.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Sorin Group USA, Inc.
Distributor
Worldwide distribution: USA (nationwide) including states of : MI and NY; and countries of: Belgium, Canada, Japan, India, Russia, United Arab Emirates.
Lot
Product Code: 60-02-50, Serial Numbers: 60S09131, 60S09133-60S09135, 60S09140-60S09142, 60S08954.
Risk classification
Class II
Area (FDA)
device
Country
United States