SCP Pump Control Panel, Product Code: 60-02-50, Serial Numbers: 60S09131, 60S09133-60S09135, 60S09140-60S09142, 60S08954. The SCP is a cardiopulmonary bypass speed control device indicated for use exclusively with the COBE Revolution pump head for speed-controlled pumping through cardiopulmonary bypass circuit for typical durations of six hours or less.
Published November 7, 2012 · Recall initiated: September 20, 2012
Summary
FDA records medical device Class II recall Z-0167-2013, published November 7, 2012 and initiated September 20, 2012. Recalling firm: Sorin Group USA, Inc.; Status: Terminated; Reported quantity: 8 units (2 units USA and 6 units OUS); Lot: Product Code: 60-02-50, Serial Numbers: 60S09131, 60S09133-60S09135, 60S09140-60S09142, 60S08954. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Stockert Centrifugal Pump (SCP) Pump Control Panel . An out of specification circuit board component within the Panel may lead to a malfunction consisting of a blank or flashing display. The flow and pressure display information may also be incorrect after the malfunction.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Sorin Group USA, Inc.
- Distributor
- Worldwide distribution: USA (nationwide) including states of : MI and NY; and countries of: Belgium, Canada, Japan, India, Russia, United Arab Emirates.
- Lot
- Product Code: 60-02-50, Serial Numbers: 60S09131, 60S09133-60S09135, 60S09140-60S09142, 60S08954.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASorin Group Aortic Arch Cannula, 7 mm x 10 in x 3/8 in, Rx Only, Sterile Product Usage: The Aortic ArchAugust 16, 2017
- FDASORIN GROUP, SMART PERFUSION PACK, SMART 1/4" PEDS PACK, REF 627262502, STERILE EO, Rx Only, 1 EAAugust 9, 2017
- FDASORIN GROUP, SMART PERFUSION PACK, SMART CIRCULATORY ASSIST PACK, REF 627255003, STERILE EO, Rx Only, 1 EAAugust 9, 2017
- FDANON-STERILE SAMPLE, SORIN GROUP, SMART PERFUSION PACK, REF 627376401, 1 EA, NOT FOR CLINICAL USEAugust 9, 2017
- FDASCP Pump Control Panel, Product Code: 60-02-15, Serial Numbers: 60S09121, 60S09123-60S09129,November 7, 2012
- FDAILLlCO® MIS Posterior Fixation System, Bone-screw Internal Spinal Fixation System, Orthosis, Spinal PedicalNovember 7, 2012
Questions about this recall
Was SCP Pump Control Panel, Product Code: 60-02-50, Serial Numbers: 60S09131, 60S09133-60S09135, 60S09140-60S09142, 60S08954. The SCP is a cardiopulmonary bypass speed control device indicated for use exclusively with the COBE Revolution pump head for speed-controlled pumping through cardiopulmonary bypass circuit for typical durations of six hours or less. recalled?
Yes. FDA published recall Z-0167-2013 on November 7, 2012. Reported hazard: Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Stockert Centrifugal Pump (SCP) Pump Control Panel . An out of specification circuit board component within the Panel may lead to a malfunction consisting of a blank or flashing display. The flow and pressure display information may also be incorrect after the malfunction.
What should I do if I have this product?
Voluntary: Firm initiated · Letter