ILLlCO® MIS Posterior Fixation System, Bone-screw Internal Spinal Fixation System, Orthosis, Spinal Pedical Fixation Orthosis, Spondylolisthesis Spinal Fixation The ILLICO MIS Posterior Fixation System is intended to facilitate the surgical correction of noncervical spinal deformities by providing temporary internal fixation and stabilization during bone graft healing and/or fusion mass development. When used for a minimally invasive posterior approach ILLlCO® MIS Instrumentation is used in conjunction with polyaxial screw components. The ILLICO MIS Posterior Fixation System is intended for use as a posterior spinal fixation device to aid in the surgical correction of various spinal deformities and pathologies of the spine. It is intended to provide stabilization during the development of fusion utilizing a bone graft. Specific indications for the ILLICO MIS Posterior Fixation System are dependent in part on the configuration of the assembled device and the method of attachment to the spine.
Published November 7, 2012 · Recall initiated: September 25, 2012
Summary
FDA records medical device Class II recall Z-0168-2013, published November 7, 2012 and initiated September 25, 2012. Recalling firm: Alphatec Spine, Inc.; Status: Terminated; Reported quantity: 209 screws; Lot: Product Codes: KW, MNI, MNH. Part NumberlLot Number: ¿PART DESCRIPTION LOT 7384S-30 Ti CANNULATED POLYAXIAL SCREW 4.5mm X 30mm 646501 73845-35 Ti CANNULATED POLYAXIAL SCREW 4.5mm X 35mm 646502 73845-40 Ti CANNULATED POLYAXIAL SCREW 4.5mm X 40mm 646503 73845-45 Ti CANNULATED POLYAXIAL SCREW 4.5mm X 45mm 646504 73855-35 Ti CANNULATED POLYAXIAL SCREW 5.5mm X 35mm 646510 73865-30 Ti CANNULATED POLYAXIAL SCREW 6.5mm X 30mm 646515 73865-35 Ti CANNULATED POLYAXIAL SCREW 6.5mm X 35mm 646516 73865-40 Ti CANNULATED POLYAXIAL SCREW 6.5mm X 40mm 646517 73865-45 Ti CANNULATED POLYAXIAL SCREW 6.5mm X 45mm 646542 73865-50 Ti CANNULATED POLYAXIAL SCREW 6.5mm X 50mm 646543 73875-35 Ti CANNULATED POLYAXIAL SCREW 7.5mm X 35mm 646546 73875-45 Ti CANNULATED POLYAXIAL SCREW 7.5mm X 45mm 646548 73875-50 Ti CANNULATED POLYAXIAL SCREW 7.5mm X 50mm 646549. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The firm initiated the recall because the stainless steel guide-wire of the Illico MIS Fixation System did not completely fit/pass through the tip of certain Illico Ti Cannulated Polyaxial Screws, therefore rendering such screws nonfunctional.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Alphatec Spine, Inc.
- Distributor
- Nationwide Distribution including AZ, IL, CA, SC, TN, TX, GA, CT, NY, FL, OK, LA, MO, NV,and UT.
- Lot
- Product Codes: KW, MNI, MNH. Part NumberlLot Number: ¿PART DESCRIPTION LOT 7384S-30 Ti CANNULATED POLYAXIAL SCREW 4.5mm X 30mm 646501 73845-35 Ti CANNULATED POLYAXIAL SCREW 4.5mm X 35mm 646502 73845-40 Ti CANNULATED POLYAXIAL SCREW 4.5mm X 40mm 646503 73845-45 Ti CANNULATED POLYAXIAL SCREW 4.5mm X 45mm 646504 73855-35 Ti CANNULATED POLYAXIAL SCREW 5.5mm X 35mm 646510 73865-30 Ti CANNULATED POLYAXIAL SCREW 6.5mm X 30mm 646515 73865-35 Ti CANNULATED POLYAXIAL SCREW 6.5mm X 35mm 646516 73865-40 Ti CANNULATED POLYAXIAL SCREW 6.5mm X 40mm 646517 73865-45 Ti CANNULATED POLYAXIAL SCREW 6.5mm X 45mm 646542 73865-50 Ti CANNULATED POLYAXIAL SCREW 6.5mm X 50mm 646543 73875-35 Ti CANNULATED POLYAXIAL SCREW 7.5mm X 35mm 646546 73875-45 Ti CANNULATED POLYAXIAL SCREW 7.5mm X 45mm 646548 73875-50 Ti CANNULATED POLYAXIAL SCREW 7.5mm X 50mm 646549
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASafeOp Patient Interface, REF: AIX11240, part of SafeOp 3: Neural Informatix SystemJuly 29, 2026
- FDAIdentiTi ALIF Standalone Interbody System: Part Number/Description 132-10-85-250 / IdentiTi ALIF SA GraftMarch 6, 2024
- FDAAtec Insignia Anterior Cervical Plate System, REF 136-0354 Insignia, ACP, 3-Level, 54 mm, Rx Only,June 16, 2021
- FDAAtec Insignia Anterior Cervical Plate System, REF 136-0248 Insignia, ACP, 2-Level, 48 mm, Rx Only,June 16, 2021
- FDASCP Pump Control Panel, Product Code: 60-02-50, Serial Numbers: 60S09131, 60S09133-60S09135,November 7, 2012
- FDAMOSAIQ MOSAIQ is an image-enabled electronic medical record system (EMR) used for oncology workflowNovember 7, 2012
Questions about this recall
Was ILLlCO® MIS Posterior Fixation System, Bone-screw Internal Spinal Fixation System, Orthosis, Spinal Pedical Fixation Orthosis, Spondylolisthesis Spinal Fixation The ILLICO MIS Posterior Fixation System is intended to facilitate the surgical correction of noncervical spinal deformities by providing temporary internal fixation and stabilization during bone graft healing and/or fusion mass development. When used for a minimally invasive posterior approach ILLlCO® MIS Instrumentation is used in conjunction with polyaxial screw components. The ILLICO MIS Posterior Fixation System is intended for use as a posterior spinal fixation device to aid in the surgical correction of various spinal deformities and pathologies of the spine. It is intended to provide stabilization during the development of fusion utilizing a bone graft. Specific indications for the ILLICO MIS Posterior Fixation System are dependent in part on the configuration of the assembled device and the method of attachment to the spine. recalled?
Yes. FDA published recall Z-0168-2013 on November 7, 2012. Reported hazard: The firm initiated the recall because the stainless steel guide-wire of the Illico MIS Fixation System did not completely fit/pass through the tip of certain Illico Ti Cannulated Polyaxial Screws, therefore rendering such screws nonfunctional.
What should I do if I have this product?
Voluntary: Firm initiated · Letter