FDAFDA · Food and drugs·Recall number Z-0183-2014

Applied Medical's Inzii® 12/15mm Retrieval System. Applieds Inzii Tissue Retrieval Bag, Model Number CD004, is classified as a Class II device. It was cleared to market via 510(k) filing K060051. The retrieval bag is a disposable specimen retrieval bag intended for use as receptacle for the collection of tissue, organs, and calculi during laparoscopic procedures. Applieds CD004 has a shelf life of three years. It is provided sterile and packaged in a heat sealed Tyvek pouch.

Published November 20, 2013 · Recall initiated: October 21, 2013

Summary

FDA records medical device Class II recall Z-0183-2014, published November 20, 2013 and initiated October 21, 2013. Recalling firm: Applied Medical Resources Corp; Status: Terminated; Reported quantity: 29,506 units; Lot: Model Number CD004. Lot Nubmers: 1161268, 1161269, 1164934, 1168358, 1168361, 1169317, 1169318, 1170694, 1170700, 1170701, 1171214, 1172482, 1172664, 1174287, 1174793, 1174857, 1175123, 1175280, 1175472, 1176555, 1177023, 1177871, 1179069, 1179392, 1179460, 1179775, 1179890, 1180291, 1180411, 1180576, 1181647, 1181648, 1181649, 1181939, 1182760, 1182934, 1183225, 1183823, 1183957, 1184273, 1184728, 1184915, 1184916, 1187779, 1188817, 1189117, 1189957, 1190294, 1190358, 1191208, 1192057, 1192969, 1193196, 1193661, 1194010, 1194666, 1194807, 1195518, 1195727, 1195997, 1197182, 1197404, 1197832, 1198055, 1198968, 1199899, 1200658, 1200828, 1200829, 1200830, 1201739, 1201740, 1201741, 1203154, 1204054. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Applied Medical is conducting a voluntary recall of the Inzii® 12/15mm retrieval system. During shipment, the retrieval system packaging has the potential to become punctured with small holes, which could compromise the sterile barrier. The likelihood of this situation to occur and result in patient harm is highly unlikely; however, out of an abundance of caution for patient safety and a commitmen

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Applied Medical Resources Corp
Distributor
Worldwide Distribution: US (nationwide) and countries of: Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Lebanon, Malaysia, Mexico, New Zealand, Netherlands, Norway, Peru, Poland, Portugal, Saudi Arabia, Singapore, Spain, South Africa, Sweden, Switzerland, United Arab Emirates, and United Kingdom.
Lot
Model Number CD004. Lot Nubmers: 1161268, 1161269, 1164934, 1168358, 1168361, 1169317, 1169318, 1170694, 1170700, 1170701, 1171214, 1172482, 1172664, 1174287, 1174793, 1174857, 1175123, 1175280, 1175472, 1176555, 1177023, 1177871, 1179069, 1179392, 1179460, 1179775, 1179890, 1180291, 1180411, 1180576, 1181647, 1181648, 1181649, 1181939, 1182760, 1182934, 1183225, 1183823, 1183957, 1184273, 1184728, 1184915, 1184916, 1187779, 1188817, 1189117, 1189957, 1190294, 1190358, 1191208, 1192057, 1192969, 1193196, 1193661, 1194010, 1194666, 1194807, 1195518, 1195727, 1195997, 1197182, 1197404, 1197832, 1198055, 1198968, 1199899, 1200658, 1200828, 1200829, 1200830, 1201739, 1201740, 1201741, 1203154, 1204054.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Applied Medical's Inzii® 12/15mm Retrieval System. Applieds Inzii Tissue Retrieval Bag, Model Number CD004, is classified as a Class II device. It was cleared to market via 510(k) filing K060051. The retrieval bag is a disposable specimen retrieval bag intended for use as receptacle for the collection of tissue, organs, and calculi during laparoscopic procedures. Applieds CD004 has a shelf life of three years. It is provided sterile and packaged in a heat sealed Tyvek pouch. recalled?

Yes. FDA published recall Z-0183-2014 on November 20, 2013. Reported hazard: Applied Medical is conducting a voluntary recall of the Inzii® 12/15mm retrieval system. During shipment, the retrieval system packaging has the potential to become punctured with small holes, which could compromise the sterile barrier. The likelihood of this situation to occur and result in patient harm is highly unlikely; however, out of an abundance of caution for patient safety and a commitmen

What should I do if I have this product?

Voluntary: Firm initiated · Letter